Abstract
This paper examines the Fisher identifiability of two coupled transport processes with substantially different transport coefficients. This work is motivated by the problem of estimating the efficacy of a novel life support technology for respiratory failure patients. The idea is to circulate an oxygen carrier through the patient’s abdomen, thereby utilizing abdominal gas diffusion for life support. The paper presents a third-order nonlinear model of the coupled dynamics of gas transport in the lungs and abdomen during this treatment. We linearize this model, reduce its order, and analyze its parameter identifiability. The main insight is that the stronger transport process in the lungs acts as a feedback mechanism that weakens the identifiability of the parameter governing the weaker abdominal transport process. Manipulating the stronger transport process through active control and/or passive design can, therefore, improve identifiability. The paper illustrates these insights using Monte Carlo simulation, showing a fourfold improvement in abdominal transport coefficient estimation accuracy through simple experimental redesign.
Funding Information
  • U.S. NSF GCR (2121110)

This publication has 29 references indexed in Scilit: