A top-loading matrix isolation apparatus for magneto-optical investigations
Authors:
- 1 August 1978
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 11 (8) , 801-804
Abstract
An apparatus is described whereby an argon matrix is deposited (together with a species of interest) by conventional matrix isolation techniques in a small, efficient, separate continuous-flow helium-cryostat and then injected in vacuo, into a large standard split-pair superconducting magnet cryostat. During injection the sample does not warm above 20K and the vacuum space around the sample is then filled with liquid helium or exchange gas. A critical example of the value of the technique is given with Zeeman anisotropy of the sharp line emission of Cl2 in argon detected at about 13600 cm-1.Keywords
This publication has 5 references indexed in Scilit:
- Magnetic circular dichroism of matrix isolated noble gas monohalidesChemical Physics Letters, 1977
- Photophysics of low lying electronic states of Cl2 in rare gas solidsThe Journal of Chemical Physics, 1976
- Laser-induced fluorescence and Raman spectra of chlorine and bromine molecules isolated in inert matricesJournal of Molecular Spectroscopy, 1975
- Magnetic circular dichroism spectroscopy of matrix-isolated speciesMolecular Physics, 1974
- Matrix Isolation Method for the Experimental Study of Unstable SpeciesThe Journal of Chemical Physics, 1954