Improved Control of Organism Concentration in Continuous Cultures of Filamentous Micro‐organisms
Authors:
- 1 April 1967
- journal article
- Published by Wiley in Journal of Applied Bacteriology
- Vol. 30 (1) , 246-250
Abstract
S ummary . Mycelium concentration and culture volume measurements in continuous cultures are affected by accretions of filamentous organisms at the overflow weir. A pinch valve is described which periodically closed the air and culture effluent tube. When released the outrush of air cleared the overflow system. With the device the standard deviations of the steady state mycelium dry weights were reduced to one third of their former value and progressive concentration of the organism was eliminated. Measurement of culture volume was corrected for gas retained in the mycelial mass.Keywords
This publication has 3 references indexed in Scilit:
- STUDIES OF THE GROWTH OF PENICILLIUM CHRYSOGENUM IN CONTINUOUS FLOW CULTURE WITH REFERENCE TO PENICILLIN PRODUCTIONJournal of Applied Bacteriology, 1960
- A TWO LITRE SCALE CONTINUOUS CULTURE APPARATUS FOR MICRO-ORGANISMSJournal of Applied Bacteriology, 1956
- Experiments with the Chemostat on Spontaneous Mutations of BacteriaProceedings of the National Academy of Sciences, 1950