Time-resolved cardiac interventional cone-beam CT reconstruction from fully truncated projections using the prior image constrained compressed sensing (PICCS) algorithm
- 5 April 2012
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 57 (9) , 2461-2476
Abstract
C-arm cone-beam CT could replace preoperative multi-detector CT scans in the cardiac interventional setting. However, cardiac gating results in view angle undersampling and the small size of the detector results in projection data truncation. These problems are incompatible with conventional tomographic reconstruction algorithms. In this paper, the prior image constrained compressed sensing (PICCS) reconstruction method was adapted to solve these issues. The performance of the proposed method was compared to that of FDK, FDK with extrapolated projection data (E-FDK), and total variation-based compressed sensing. A canine projection dataset acquired using a clinical C-arm imaging system supplied realistic cardiac motion and anatomy for this evaluation. Three different levels of truncation were simulated. The relative root mean squared error and the universal image quality index were used to quantify the reconstruction accuracy. Three main conclusions were reached. (1) The adapted version of the PICCS algorithm offered the highest image quality and reconstruction accuracy. (2) No meaningful variation in performance was observed when the amount of truncation was changed. (3) This study showed evidence that accurate interior tomography with an undersampled acquisition is possible for realistic objects if a prior image with minimal artifacts is available.Keywords
This publication has 66 references indexed in Scilit:
- Prior image constrained compressed sensing: Implementation and performance evaluationMedical Physics, 2011
- Evaluation of sparse-view reconstruction from flat-panel-detector cone-beam CTPhysics in Medicine & Biology, 2010
- High-order total variation minimization for interior tomographyInverse Problems, 2010
- Performance comparison between total variation (TV)-based compressed sensing and statistical iterative reconstruction algorithmsPhysics in Medicine & Biology, 2009
- Supplemental analysis on compressed sensing based interior tomographyPhysics in Medicine & Biology, 2009
- Compressed sensing based interior tomographyPhysics in Medicine & Biology, 2009
- High temporal resolution and streak-free four-dimensional cone-beam computed tomographyPhysics in Medicine & Biology, 2008
- Image reconstruction in circular cone-beam computed tomography by constrained, total-variation minimizationPhysics in Medicine & Biology, 2008
- Exact Interior Reconstruction with Cone‐Beam CTInternational Journal of Biomedical Imaging, 2007
- A General Local Reconstruction Approach Based on a Truncated Hilbert TransformInternational Journal of Biomedical Imaging, 2007