Coincidence Summing Corrections in Gamma-Ray Spectrometry Using GEANT4 Code
- 16 June 2009
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Nuclear Science
- Vol. 56 (3) , 1531-1536
Abstract
In this paper coincidence summing corrections in gamma-ray spectrometry have been analyzed in depth. Experimental setup included a n-type germanium detector and two efficient geometries of measurement (Marinelli beaker and air filter) calibrated with multi gamma radionuclides in the energy range of 40-1500 keV. Monte Carlo simulations were carried out using the GEANT4 code in order to develop this work. Firstly an optimization of the detector dimensions was performed to obtain the full-energy peak efficiency curve. Next, the simulation of the decay schemes for involved multi gamma radionuclides were appended to the GEANT4 code including beta, alpha, EC decay, together with the subsequent relaxation process consisting of a cascade of fluorescent X-rays and Auger electrons. In order to test the simulations an comparison between an experimental and a calculated spectra of 133Ba point source was carried out. Finally coincidence summing effects for gamma-ray emission and X-ray, Auger electron, beta and EC electron were studied in detail.Keywords
This publication has 18 references indexed in Scilit:
- Validation of Geant4 X-ray fluorescence transitions - validation of Geant4 electromagnetic models against calorimetry measurements in the energy range up to 1 MeV2008
- Validation of Geant4 Atomic Relaxation Against the NIST Physical Reference DataIEEE Transactions on Nuclear Science, 2007
- Geant4 Atomic RelaxationIEEE Transactions on Nuclear Science, 2007
- A revision of energy and resolution calibration method of Ge detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2006
- Geant4 developments and applicationsIEEE Transactions on Nuclear Science, 2006
- GEANT4 code for simulation of a germanium gamma-ray detector and its application to efficiency calibrationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2004
- Efficiency calculation and coincidence summing correction for germanium detectors by Monte Carlo simulationIEEE Transactions on Nuclear Science, 2002
- Geant4-a new Monte Carlo toolkit for simulating space radiation shielding and effects2002
- Coincidence summing in gamma and X-ray spectrometryNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1993
- Coincidence-summing corrections for extended sources in gamma-ray spectrometry using Monte Carlo simulationNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1992