Development of pressure-driven micro active catheter using membrane micro emboss following excimer laser ablation (MeME-X) process
Authors:
- 1 May 2009
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 10504729,p. 4469-4472
Abstract
The world's smallest micro active catheter of φ∼300µm was successfully fabricated using membrane micro emboss following excimer laser ablation (MeME-X) process. The catheter has a one-sided hollow bellows at the tip made of thin polymer membrane. The bellows is composed of a series of folded micro-chambers and microchannels connecting the micro-chambers. The folded-chambers expand on one side by increasing inner water pressure using a syringe, thus the whole bellows bends toward one direction within 0 to 180 degree. This micro active catheter should be useful for safe intravascular surgery in narrow and complicated blood vessels. Moreover, the nonelectrical actuation mechanism of this catheter can be widely applied to micro-actuators and micro-robots which need high safety.Keywords
This publication has 6 references indexed in Scilit:
- A new approach to develop ionic polymer–metal composites (IPMC) actuator: Fabrication and control for active catheter systemsSensors and Actuators A: Physical, 2007
- Artificial Capillary Network Chip for in Vitro 3D Tissue Culture2007
- Pneumatic micro finger as endeffecter of robot2006
- Mechanisms of Phrenic Nerve Injury During Radiofrequency Ablation at the Pulmonary Vein OrificeJournal of Cardiovascular Electrophysiology, 2005
- An Unusual Case of Pulmonary Artery Catheter MalfunctionAnesthesia & Analgesia, 2002
- An active guide wire with shape memory alloy bending actuator fabricated by room temperature processSensors and Actuators A: Physical, 2002