Difference in normal limit values of nerve conduction parameters between Westerners and Japanese people might need to be considered when diagnosing diabetic polyneuropathy using a Point‐of‐Care Sural Nerve Conduction Device (NC‐stat®/DPNCheck™)
Authors:
Open Access
- 6 March 2018
- journal article
- research article
- Published by Wiley in Journal of Diabetes Investigation
- Vol. 9 (5) , 1173-1181
Abstract
Aim/Introduction: Studies on a novel point‐of‐care device for nerve conduction study called DPNCheck have been limited to Westerners. We aimed to clarify Japanese normal limits of nerve action potential amplitude (Amp) and conduction velocity by DPNCheck (investigation I), and the validity of DPNCheck to identify diabetic symmetric sensorimotor polyneuropathy (DSPN; investigation II).Materials and Methods: For investigation I, 463 non‐neuropathic Japanese participants underwent DPNCheck examinations. Regression formulas calculating the normal limits of Amp and conduction velocity (Japanese regression formulas [JRF]) were determined by quantile regression and then compared with regression formulas of individuals from the USA (USRF). For investigation II, in 92 Japanese diabetes patients, ‘probable DSPN’ was diagnosed and nerve conduction abnormalities (NCA1: one or more abnormalities, and NCA2: two abnormalities in Amp and conduction velocity) were determined. Validity of NCAs to identify ‘probable DSPN’ was evaluated by determining sensitivity, specificity, reproducibility (kappa‐coefficient) and the area under the curve of receiver operating characteristic curves.Results: For investigation I, JRF was different from USRF, and normal limits by JRF were higher than that of USRF. The prevalence of Amp abnormality calculated by JRF was significantly higher than that of USRF. For investigation II, the sensitivity, specificity and reproducibility of NCA1 and NCA2 judged from JRF were 85%, 86% and 0.57, and 43%, 100% and 0.56, respectively. These values of JRF were higher than those of USRF. The area under the curve of JRF (0.89) was larger than USRF (0.82).Conclusions: A significant difference in the normal limits of nerve conduction parameters by DPNCheck between Japanese and USA individuals was suggested. Validity to identify DSPN of NCAs might improve by changing the judgment criteria from USRF to JRF.Keywords
Funding Information
- Japan Society for the Promotion of Science (JP15K01723)
This publication has 11 references indexed in Scilit:
- Association of nerve conduction impairment and insulin resistance in children with obesityChild's Nervous System, 2016
- Quantitative assessment of chemotherapy‐induced peripheral neurotoxicity using a point‐of‐care nerve conduction deviceCancer Science, 2016
- Reliability and Validity of a Point-of-Care Sural Nerve Conduction Device for Identification of Diabetic NeuropathyPLOS ONE, 2014
- Pro198Leu missense polymorphism of the glutathione peroxidase 1 gene might be a common genetic predisposition of distal symmetric polyneuropathy and macrovascular disease in Japanese type 2 diabetic patientsJournal of Diabetes Investigation, 2011
- Diabetic Neuropathies: Update on Definitions, Diagnostic Criteria, Estimation of Severity, and TreatmentsDiabetes Care, 2010
- Developing reference data for nerve conduction studies: An application of quantile regressionMuscle & Nerve, 2009
- Multi-Site Testing With a Point-of-Care Nerve Conduction Device Can Be Used in an Algorithm to Diagnose Diabetic Sensorimotor PolyneuropathyDiabetes Care, 2008
- Validation of a Novel Point-of-Care Nerve Conduction Device for the Detection of Diabetic Sensorimotor PolyneuropathyDiabetes Care, 2006
- Sural nerve conduction parameters in normal subjects related to age, gender, temperature, and height: A reappraisalMuscle & Nerve, 1992
- Influence of Height on Quantitative Sensory, Nerve-Conduction, and Clinical Indices of Diabetic Peripheral NeuropathyDiabetes Care, 1987