Automation for clinical CD4 T‐cell enumeration, a desirable tool in the hands of skilled operators
Authors:
- 15 March 2016
- journal article
- review article
- Published by Wiley in Cytometry Part B: Clinical Cytometry
- Vol. 92 (6) , 445-450
Abstract
Background: Automation in HIV clinical flow cytometry when appropriately applied brings considerable standardisation benefits. The Canadian Immunology Quality Assessment Program (CIQAP) detected situations where operators did not manually override automated software in the event of improper output on the Epics XL and FC500 CD4 immunophenotyping platforms. The automated gating algorithm identifies lymphocytes using a double gate strategy based on CD45 × side scatter (SS) gating and a light scatter FS × SS gate known to fail with sub optimal specimens.Method: To generate correct interpretation and results CIQAP introduced a simple protocol modification, bypassing the light scatter gate to include all cells characterized by the CD45 gate. Seventeen problem cases were reanalysed for both absolute and relative T‐cell subsets accuracy and compared to the CIQAP group mean values. Results were found to be associated with the percentage of lymphocytes excluded by the automated light scatter gate.Results: The modified manual protocol resolved poor performance in 14 instances out of 17 problem cases. It was found to improve accuracy when the light scatter gate excluded greater than 5% of the cells. The remaining three cases had a lymphocyte recovery of greater than 94.6% in the original automated analysis.Conclusion: There is a risk in relying solely on automated gating procedures when using the Epics XL and FC500 CD4 immunophenotyping platforms. Laboratory managers have the responsibility to intervene when required. EQA providers are equally responsible to alert the clinical laboratories of the need to update operator training to deal with stressed specimens. © 2016 International Clinical Cytometry SocietyKeywords
This publication has 6 references indexed in Scilit:
- Compatibility of Stabilized Whole Blood Products with CD4 Technologies and Their Suitability for Quality Assessment ProgramsPLOS ONE, 2014
- Flow Cytometry: A Multipurpose Technology for a Wide Spectrum of Global Biosecurity ApplicationsSLAS Technology, 2009
- Selection of lymphocyte gating protocol has an impact on the level of reliability of T‐cell subsets in aging specimensCytometry, 2002
- Use of CD45 gating in three and four‐color flow cytometric immunophenotyping: Guideline from the national institute of allergy and infectious diseases, division of AIDSCytometry, 2002
- Adoption of Single-Platform Technologies for Enumeration of Absolute T-Lymphocyte Subsets in Peripheral BloodClinical and Diagnostic Laboratory Immunology, 2000
- Use of CD45 fluorescence and side-scatter characteristics for gating lymphocytes when using the whole blood lysis procedure and flow cytometryCytometry, 1996