Post marketing study of hemodynamic and hematological noninvasive readings in a blood bank
Open Access
- 1 January 2018
- journal article
- research article
- Published by SAGE Publications in Sage Open Medicine
- Vol. 6
Abstract
This validation test was conducted in the Fujisan Blood Bank, Fortaleza, Brazil and evaluated the noninvasive TensorTip MTX (MTX, Cnoga Medical Ltd.) readings of hemoglobin, hematocrit, red blood cells, blood pressure, and heart rate compared to reference lab device readings. Generally, these parameters are measured from venous or capillary blood samples run on a laboratory analyzer or handheld invasive testing devices. Needle sticks are inconvenient to blood donors with relatively high exposure risks. To our vision, noninvasive determination would be of benefit to blood contributors and medical professional teams; it would be fast and painless. A total of 334 subjects were included in the Fujisan blood bank validation (65% male, 35% female). Hemoglobin, hematocrit, and red blood cells, as well as blood pressure and heart rate, were measured noninvasively using the MTX device and were compared to venous blood samples run on two laboratory hematology analyzers (Horiba ABX Micros60 and Siemens blood count analyzer), to digital sphygmomanometer (OMRON BP786) and to manual auscultation. The noninvasive measurement with the appropriate virtual arm cuff setting was performed simultaneously with the blood sample extraction of the reference devices measurement. There was no statistically significant difference (p > 0.05, paired, two-tailed t-test) between the average daily hemoglobin, hematocrit, and red blood cells measurements provided by the MTX device and the laboratory hematology analyzer. In addition, there was no significant difference between the daily blood pressure and heart rate results provided by the MTX device and the digital and manual sphygmomanometers. The error calculated between the MTX and the reference device was found to be sufficiently accurate according to the relevant standards. The MTX accuracy of noninvasive hemoglobin, hematocrit, red blood cells, blood pressure, and heart rate measurements satisfies the industrial standards; therefore, the device enables more accurate, efficient, and effective patient care.Keywords
This publication has 17 references indexed in Scilit:
- Comparison of the Accuracy of Noninvasive Hemoglobin Sensor (NBM-200) and Portable Hemoglobinometer (HemoCue) with an Automated Hematology Analyzer (LH500) in Blood Donor ScreeningAnnals of Laboratory Medicine, 2013
- Accuracy of Noninvasive and Continuous Hemoglobin Measurement by Pulse Co-Oximetry During Preoperative PhlebotomyJournal of Intensive Care Medicine, 2013
- Noninvasive methods for haemoglobin screening in prospective blood donorsVox Sanguinis, 2013
- Student’s t-Test2013
- Non-Invasive Measurement of Hemoglobin: Assessment of Two Different Point-of-Care TechnologiesPLOS ONE, 2012
- Accuracy of a continuous noninvasive hemoglobin monitor in intensive care unit patients*Critical Care Medicine, 2011
- The Accuracy of Noninvasive and Continuous Total Hemoglobin Measurement by Pulse CO-Oximetry in Human Subjects Undergoing HemodilutionAnesthesia & Analgesia, 2010
- Pulse oximetry: its invention, theory, and futureJournal of Anesthesia, 2003
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- A New Biometrical Procedure for Testing the Equality of Measurements from Two Different Analytical Methods. Application of linear regression procedures for method comparison studies in Clinical Chemistry, Part IClinical Chemistry and Laboratory Medicine (CCLM), 1983