Relative motion of hip stems under load. An in vitro study of symmetrical, asymmetrical, and custom asymmetrical designs.
Authors:
- P S WalkerPWP S Walker
- 1 January 1994
- journal article
- Published by Wolters Kluwer Health in Journal of Bone and Joint Surgery
- Vol. 76 (1) , 95-103
Abstract
Ed to simulate the forces acting on a hip joint in a single-limb stance. A force of 1000 newtons was applied three times in order to seat the prosthesis, and then a cyclical force of 1000 newtons was applied to the head of the prosthesis; readings of relative motion were taken every 500 cycles, until 2500 cycles had been reached. The relative motion diminished as the number of cycles increased, with most of the reduction occurring within the first 500 cycles. The symmetrical stem had the least relative axial motion, with an average of six micrometers, but the other stems also had low values: eleven micrometers for the custom stem, and nineteen micrometers for the asymmetrical stem. In terms of axial rotation, however, the symmetrical stem had several times more motion than did the other two stems. Motion of the stem tip toward or away from the canal in the frontal plane and in the sagittal plane also was higher for the symmetrical stem. Axial migration (permanent sinkage after 2500 cycles) was very similar between all three types of stem. In a torsional loading test, the symmetrical stem showed about ten times more relative rotational motion than did the other two stems. Over-all, the custom stem and the asymmetrical stem had similar results.(ABSTRACT TRUNCATED AT 250 WORDS) Symmetrical, asymmetrical, and custom asymmetrical stems without collars were manufactured and inserted without cement, in that order, into each of eight femora. Displacement transducers were mounted at different locations to measure relative motion between the stem and the bone. A load was applied to simulate the forces acting on a hip joint in a single-limb stance. A force of 1000 newtons was applied three times in order to seat the prosthesis, and then a cyclical force of 1000 newtons was applied to the head of the prosthesis; readings of relative motion were taken every 500 cycles, until 2500 cycles had been reached. The relative motion diminished as the number of cycles increased, with most of the reduction occurring within the first 500 cycles. The symmetrical stem had the least relative axial motion, with an average of six micrometers, but the other stems also had low values: eleven micrometers for the custom stem, and nineteen micrometers for the asymmetrical stem. In terms of axial rotation, however, the symmetrical stem had several times more motion than did the other two stems. Motion of the stem tip toward or away from the canal in the frontal plane and in the sagittal plane also was higher for the symmetrical stem. Axial migration (permanent sinkage after 2500 cycles) was very similar between all three types of stem. In a torsional loading test, the symmetrical stem showed about ten times more relative rotational motion than did the other two stems. Over-all, the custom stem and the asymmetrical stem had similar results.(ABSTRACT TRUNCATED AT 250 WORDS) Copyright © 1994 by The Journal of Bone and Joint Surgery, Incorporated...Keywords
This publication has 28 references indexed in Scilit:
- A computer-based biomechanical analysis of the three-dimensional motion of cementless hip prosthesesJournal of Biomechanics, 1992
- Telemeterized in vivo hip joint force data: A report on two patients after total hip surgeryJournal of Orthopaedic Research, 1991
- Comparison of the stability of press‐fit hip prosthesis femoral stems using a synthetic model femurJournal of Orthopaedic Research, 1991
- Micromotion of cemented and uncemented femoral componentsThe Journal of Bone and Joint Surgery. British volume, 1991
- In vitro analysis of proximal femoral strains using PCA femoral implants and a hip-abductor muscle simulatorThe Journal of Arthroplasty, 1989
- Abductor function after total hip replacement. An electromyographic and clinical reviewThe Journal of Bone and Joint Surgery. British volume, 1989
- THE PREDICTION OF FORCES IN JOINT STRUCTURESExercise and Sport Sciences Reviews, 1981
- Load actions transmitted by implantsJournal of Biomedical Engineering, 1979
- In vivo records of hip loads using a femoral implant with telemetric output (a prelimary report)Journal of Biomedical Engineering, 1979
- Effect of embalming on the mechanical properties of beef boneJournal of Applied Physiology, 1964