The appearance and effects of metallic implants in CT images
- 23 May 2013
- journal article
- research article
- Published by Springer Nature in Australasian Physics & Engineering Sciences in Medicine
- Vol. 36 (2) , 209-217
Abstract
The computed tomography (CT) imaging artefacts that metallic medical implants produce in surrounding tissues are usually contoured and over-ridden during radiotherapy treatment planning. In cases where radiotherapy treatment beams unavoidably pass though implants, it is especially important to understand the imaging artefacts that may occur within the implants themselves. This study examines CT images of a set of simple metallic objects, immersed in water, in order to evaluate reliability and variability of CT numbers (Hounsfield units, HUs) within medical implants. Model implants with a range of sizes (heights from 2.2 to 49.6 mm), electron densities (from 2.3 to 7.7 times the electron density of water) and effective atomic numbers (from 3.9 to 9.0 times the effective atomic number of water in a CT X-ray beam) were created by stacking metal coins from several currencies. These ‘implants’ were CT scanned within a large (31.0 cm across) and a small (12.8 cm across) water phantom. Resulting HU values are as much as 50 % lower than the result of extrapolating standard electron density calibration data (obtained for tissue and bone densities) up to the metal densities and there is a 6 % difference between the results obtained by scanning with 120 and 140 kVp tube potentials. Profiles through the implants show localised cupping artefacts, within the implants, as well as a gradual decline in HU outside the implants that can cause the implants’ sizes to be over estimated by 1.3–9.0 mm. These effects are exacerbated when the implants are scanned in the small phantom or at the side of the large phantom, due to reduced pre-hardening of the X-ray beam in these configurations. These results demonstrate the necessity of over-riding the densities of metallic implants, as well as their artefacts in tissue, in order to obtain accurate radiotherapy dose calculations.Keywords
This publication has 22 references indexed in Scilit:
- Dosimetric effects of a high-density spinal implantJournal of Physics: Conference Series, 2013
- Metal artifact reduction by dual energy computed tomography using monoenergetic extrapolationEuropean Radiology, 2011
- THE DEVELOPMENT OF A MONTE CARLO SYSTEM TO VERIFY RADIOTHERAPY TREATMENT DOSE CALCULATIONSRadiotherapy and Oncology, 2009
- Effect of dental restorations and prostheses on radiotherapy dose distribution: a Monte Carlo studyJournal of Applied Clinical Medical Physics, 2009
- Monte Carlo characterization of materials for prosthetic implants and dosimetric validation of Pinnacle3 TPSNuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms, 2008
- Technical approaches to the optimisation of CTPhysica Medica, 2008
- Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculationsPhysics in Medicine & Biology, 2008
- Artifacts in CT: Recognition and AvoidanceRadioGraphics, 2004
- Radiotherapy dose calculations in the presence of hip prosthesesMedical Dosimetry, 2003
- Prevalence of total hip replacement: how much demand has been met?Journal of Epidemiology and Community Health, 1994