Test–retest, inter- and intra-rater reproducibility of size measurements of focal bone marrow lesions in MRI in patients with multiple myeloma
Authors:
- Markus WennmannMarkus Wennmann
- Martin GrözingerMGMartin Grözinger
- Vivienn WeruVWVivienn Weru
- Thomas HielscherTHThomas Hielscher
- Lukas Thomas RotkopfLRLukas Thomas Rotkopf
- Fabian BauerFBFabian Bauer
- Regula GnirsRGRegula Gnirs
- Tobias NonnenmacherTNTobias Nonnenmacher
- Sandra SauerSSSandra Sauer
- Hartmut GoldschmidtHGHartmut Goldschmidt
- Niels WeinholdNWNiels Weinhold
- David BonekampDBDavid Bonekamp
Open Access
- 12 April 2023
- journal article
- research article
- Published by Oxford University Press (OUP) in British Journal of Radiology
- Vol. 96 (1145) , 20220745
Abstract
Objective: To investigate the reproducibility of size measurements of focal bone marrow lesions (FL) in MRI in patients with monoclonal plasma cell disorders under variation of patient positioning and observer. Methods: A data set from a prospective test–retest study was used, in which 37 patients with a total of 140 FL had undergone 2 MRI scans with identical parameters after patient repositioning. Two readers measured long and short axis diameter on the initial scan in T 1 weighted, T 2 weighted short tau inversion recovery and diffusion-weighted imaging sequences. The first reader additionally measured FL on the retest-scan. The Bland–Altman method was used to assess limits of agreement (LoA), and the frequencies of absolute size changes were calculated. Results: In the simple test–retest experiment with one identical reader, a deviation of ≥1 mm / ≥2 mm / ≥3 mm for the long axis diameter in T 1 weighted images was observed in 66% / 25% / 8% of cases. When comparing measurements of one reader on the first scan to the measurement of the other reader on the retest scan, a change of ≥1 mm / ≥3 mm / ≥5 mm for the long axis diameter in T 1 weighted images was observed in 78% / 21% / 5% of cases. Conclusion: Small deviations in FL size are common and probably due to variation in patient positioning or inter-rater variability alone, without any actual biological change of the FL. Knowledge of the uncertainty associated with size measurements of FLs is critical for radiologists and oncologists when interpreting changes in FL size in clinical practice and in clinical trials. Advances in knowledge: According to the MY-RADs criteria, size measurements of focal lesions in MRI are now of relevance for response assessment in patients with monoclonal plasma cell disorders. Size changes of 1 or 2 mm are frequently observed due to uncertainty of the measurement only, while the actual focal lesion has not undergone any biological change. Size changes of at least 6 mm or more in T 1 weighted or T 2 weighted short tau inversion recovery sequences occur in only 5% or less of cases when the focal lesion has not undergone any biological change.Keywords
This publication has 40 references indexed in Scilit:
- International Myeloma Working Group updated criteria for the diagnosis of multiple myelomaThe Lancet Oncology, 2014
- Monitoring the response of bone metastases to treatment with Magnetic Resonance Imaging and nuclear medicine techniques: A review and position statement by the European Organisation for Research and Treatment of Cancer imaging groupEuropean Journal of Cancer, 2014
- The diagnostic value of SE MRI and DWI of the spine in patients with monoclonal gammopathy of undetermined significance, smouldering myeloma and multiple myelomaEuropean Radiology, 2014
- Whole-Body Diffusion-weighted MR Imaging for Assessment of Treatment Response in MyelomaRadiology, 2014
- Predictive value of longitudinal whole-body magnetic resonance imaging in patients with smoldering multiple myelomaLeukemia, 2014
- Bone metastases from prostate, breast and multiple myeloma: differences in lesion conspicuity at short-tau inversion recovery and diffusion-weighted MRIBritish Journal of Radiology, 2012
- Variability of Lung Tumor Measurements on Repeat Computed Tomography Scans Taken Within 15 MinutesJournal of Clinical Oncology, 2011
- Evaluating Variability in Tumor Measurements from Same-day Repeat CT Scans of Patients with Non–Small Cell Lung CancerRadiology, 2009
- New response evaluation criteria in solid tumours: Revised RECIST guideline (version 1.1)European Journal of Cancer, 2009
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986