Tolerance levels of mass density for CT number calibration in photon radiation therapy
Open Access
- 13 May 2019
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 20 (6) , 45-52
Abstract
Computed tomography (CT) data are required to calculate the dose distribution in a patient’s body. Generally, there are two CT number calibration methods for commercial radiotherapy treatment planning system (RTPS), namely CT number‐relative electron density calibration (CT‐RED calibration) and CT number‐mass density calibration (CT‐MD calibration). In a previous study, the tolerance levels of CT‐RED calibration were established for each tissue type. The tolerance levels were established when the relative dose error to local dose reached 2%. However, the tolerance levels of CT‐MD calibration are not established yet. We established the tolerance levels of CT‐MD calibration based on the tolerance levels of CT‐RED calibration. In order to convert mass density (MD) to relative electron density (RED), the conversion factors were determined with adult reference computational phantom data available in the International Commission on Radiological Protection publication 110 (ICRP‐110). In order to validate the practicability of the conversion factor, the relative dose error and the dose linearity were validated with multiple RTPSes and dose calculation algorithms for two groups, namely, CT‐RED calibration and CT‐MD calibration. The tolerance levels of CT‐MD calibration were determined from the tolerance levels of CT‐RED calibration with conversion factors. The converted RED from MD was compared with actual RED calculated from ICRP‐110. The conversion error was within ±0.01 for most standard organs. It was assumed that the conversion error was sufficiently small. The relative dose error difference for two groups was less than 0.3% for each tissue type. Therefore, the tolerance levels for CT‐MD calibration were determined from the tolerance levels of CT‐RED calibration with the conversion factors. The MD tolerance levels for lung, adipose/muscle, and cartilage/spongy‐bone corresponded to ±0.044, ±0.022, and ±0.045 g/cm3, respectively. The tolerance levels were useful in terms of approving the CT‐MD calibration table for clinical use.Keywords
This publication has 11 references indexed in Scilit:
- Assessment of the variation in CT scanner performance (image quality and Hounsfield units) with scan parameters, for image optimisation in radiotherapy treatment planningPhysica Medica, 2018
- Tolerance levels of CT number to electron density table for photon beam in radiotherapy treatment planning systemJournal of Applied Clinical Medical Physics, 2017
- Accuracy of dose calculations on kV cone beam CT images of lung cancer patientsMedical Physics, 2016
- Modeling of body tissues for Monte Carlo simulation of radiotherapy treatments planned with conventional x-ray CT systemsPhysics in Medicine & Biology, 2016
- Technical Note: Dosimetric evaluation of Monte Carlo algorithm in iPlan for stereotactic ablative body radiotherapy (SABR) for lung cancer patients using RTOG 0813 parametersJournal of Applied Clinical Medical Physics, 2015
- Clinical introduction of Monte Carlo treatment planning for lung stereotactic body radiotherapyJournal of Applied Clinical Medical Physics, 2014
- Investigation of the usability of conebeam CT data sets for dose calculationRadiation Oncology, 2008
- Quality assurance for computed-tomography simulators and the computed-tomography-simulation process: Report of the AAPM Radiation Therapy Committee Task Group No. 66Medical Physics, 2003
- Tolerance levels for quality assurance of electron density values generated from CT in radiotherapy treatment planningPhysics in Medicine & Biology, 2002
- An electron density calibration phantom for CT-based treatment planning computersMedical Physics, 1992