Discovery of New Blood Pressure Phenotypes and Relation to Accuracy of Cuff Devices Used in Daily Clinical Practice
- 1 June 2018
- journal article
- research article
- Published by Wolters Kluwer Health in Hypertension
- Vol. 71 (6) , 1239-1247
Abstract
Cuff blood pressure (BP) is the reference standard for management of high BP, but the method is inaccurate and can lead to BP misclassification. The aims of this study were to determine whether distinctive BP phenotypes exist based on BP transmission (amplification) variability from central-to-peripheral arteries and whether applying one standard cuff BP measurement approach (eg, oscillometry) to all people could discriminate the BP phenotypes. Intra-arterial BP was measured at the ascending aorta and brachial and radial arteries in 126 participants (61±10 years; 69% male) after coronary angiography. Central-to-peripheral systolic BP (SBP) transmission (SBP amplification) was defined by ≥5 mm Hg SBP increase between the aorta-to-brachial or brachial-to-radial arteries. Standard cuff BP was measured 4 different times using 3 different devices. Three independent investigators also provided data (n=255 from 4 studies) using another 3 separate cuff BP devices. Four distinct BP phenotypes were discovered based on variability in SBP amplification: phenotype 1, both aortic-to-brachial and brachial-to-radial SBP amplification; phenotype 2, only aortic-to-brachial SBP amplification; phenotype 3, only brachial-to-radial SBP amplification; and phenotype 4, neither aortic-to-brachial nor brachial-to-radial SBP amplification. Aortic SBP was significantly higher among phenotypes 3 and 4 compared with phenotypes 1 and 2 ( P =0.00074), but this was not discriminated using any standard cuff BP measures ( P =0.31). Data from independent investigators confirmed the key findings. This is the first-in-human discovery of BP phenotypes that have significantly different BPs, but which are not discriminated by standard cuff BP devices used in daily clinical practice. Improved BP device accuracy may be achieved by considering individual phenotypic BP differences.Keywords
This publication has 35 references indexed in Scilit:
- Oscillometric sphygmomanometersBlood Pressure Monitoring, 2012
- Estimation of central aortic systolic pressure from the second systolic peak of the peripheral upper limb pulse depends on central aortic pressure waveform morphologyJournal Of Hypertension, 2012
- Age-Related Changes in Aortic Arch GeometryJACC, 2011
- Role of Pulse Pressure Amplification in Arterial HypertensionHypertension, 2009
- Wave-energy patterns in carotid, brachial, and radial arteries: a noninvasive approach using wave-intensity analysisAmerican Journal of Physiology-Heart and Circulatory Physiology, 2005
- Recommendations for Blood Pressure Measurement in Humans and Experimental AnimalsHypertension, 2005
- Direct Blood Pressure Measurement —Dynamic Response RequirementsAnesthesiology, 1981
- Aortic input impedance in normal man: relationship to pressure wave forms.Circulation, 1980
- Systolic versus diastolic blood pressure and risk of coronary heart diseaseThe American Journal of Cardiology, 1971
- Recommendations for Human Blood Pressure Determinations by SphygmomanometersCirculation, 1951