Soft stereolithographic 3D printed phantoms for dual-modality particle image velocimetry (PIV)
Open Access
- 8 January 2025
- journal article
- research article
- Published by Springer Nature in Experiments in Fluids
- Vol. 66 (1) , 1-15
Abstract
The fabrication of arterial flow phantoms for fluid dynamics studies suitable for particle image velocimetry (PIV) techniques has presented challenges. Current 3D-printed blood flow phantoms with suitable transparency for optical PIV (laserPIV) are restricted to rigid materials far from those of arterial properties. Conversely, while soft 3D-printed phantoms demonstrate promise for sufficient acoustical transparency for ultrasound PIV (echoPIV), their optical translucency presents challenges for laserPIV applicability. This dual-modality approach leverages the high spatial resolution of laserPIV for in-vitro applications and the ability of echoPIV to quantify flow in both in-vivo and in-vitro application (also inside stents), providing a more comprehensive understanding of flow dynamics. In this study, we present a series of coated thin-walled 3D-printed compliant phantoms suitable for dual-modality PIV flow imaging (i.e., laserPIV and echoPIV) methods, overcoming current 3D-printable material limitations. Stereolithographic (SLA) 3D printing was used to fabricate pipe flow phantoms from a set of commercial soft resins (flexible and elastic) as vascular tissue surrogates. To overcome low transparency and poor surface finish of soft resins, we coated the 3D-printed flow phantoms with a soft, optically transparent, photo-activated polymeric coating. The feasibility of performing dual-modality PIV was tested in an in-vitro flow setup. Our results show that the average normalized root mean square errors obtained from comparing laserPIV and echoPIV velocity profiles against the analytical solutions were 3.2% and 5.1%, and 3.3% and 5.3% for the flexible and elastic phantoms, respectively. These results indicate that dual-modality PIV flow imaging is feasible in the 3D-printed coated phantoms, promoting its future use in fabricating clinically-relevant flow phantoms. Graphical abstractKeywords
This publication has 65 references indexed in Scilit:
- Design, manufacturing, and multi‐modal imaging of stereolithography 3D printed flexible intracranial aneurysm phantomsMedical Physics, 2024
- Fabrication of Soft Transparent Patient‐Specific Vascular Models with Stereolithographic 3D printing and Thiol‐Based Photopolymerizable CoatingsMacromolecular Rapid Communications, 2024
- The Supera Interwoven Nitinol Stent as a Flow Diverting Device in Popliteal AneurysmsCardioVascular and Interventional Radiology, 2022
- Validation study of computational fluid dynamics models of hemodynamics in the human aortaProceedings in Applied Mathematics and Mechanics, 2019
- Comparison of in vitro flows past a mechanical heart valve in anatomical and axisymmetric aorta modelsExperiments in Fluids, 2017
- On the rheology of refractive-index-matched, non-Newtonian blood-analog fluids for PIV experimentsExperiments in Fluids, 2016
- Particle image velocimetry in viscoelastic fluids and particle interaction effectsExperiments in Fluids, 2013
- Time-resolved PIV measurements of the flow field in a stenosed, compliant arterial modelExperiments in Fluids, 2013
- Particle image velocimetry in a foam-like porous structure using refractive index matching: a method to characterize the hydrodynamic performance of porous structuresExperiments in Fluids, 2012
- Refractive index matching methods for liquid flow investigationsExperiments in Fluids, 1994