Turbulent/Viscous Interactions Control Doppler/Catheter Pressure Discrepancies in Aortic Stenosis
Authors:
- 1 December 1996
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation
- Vol. 94 (11) , 2975-2981
Abstract
Background Despite good correlation between Doppler and catheter pressure drops in numerous reports, it is well known that Doppler tends to apparently overestimate pressure drops obtained by cardiac catheterization. Neither (1) simplification of the Bernoulli equation nor (2) pressure recovery effects can explain this dilemma when taken alone. This study addressed the hypothesis that a Reynolds number–based approach, which characterizes (1) and (2), provides a first step toward better agreement of catheter and Doppler assessments of pressure drops. Methods and Results Doppler and catheter pressure drops were studied in an in vitro model designed to isolate the proposed Reynolds number effect and in a sheep model with varying degrees of stenosis. Doppler pressure drops in vitro correlated with the directly measured pressure drop for individual valves ( r =.935, .960, .985, .984, .989, and .975) but with markedly different slopes and intercepts. A Bland-Altman type plot showed no useful pattern of discrepancy. The Reynolds number was successful in collapsing the data into the profile proposed in the hypothesis. Parallel results were found in the animal model. Conclusions Apparent overestimation of net pressure drop by Doppler is due to pressure recovery effects, and these effects are countered by both viscous effects and inertial/turbulent effects. Only by reconciliation of discrepancies by use of a quantity such as Reynolds number that embodies the relative importance of competing factors can the noninvasive and invasive methods be connected. This study shows that a Reynolds number–based approach accomplishes this goal both in the idealized in vitro setting and in a biological system.Keywords
This publication has 9 references indexed in Scilit:
- Effect of stenosis geometry on the Doppler-catheter gradient relation in vitro: A manifestation of pressure recoveryJACC, 1993
- Pressure recovery in aortic stenosis: An in vitro study in a pulsatile flow modelJACC, 1992
- Doppler evaluation of aortic valve area in children with aortic stenosisJACC, 1991
- Pressure recovery distal to a stenosis: Potential cause of gradient “verestimation” by Doppler echocardiographyJACC, 1989
- Intraoperative doppler echocardiography in hypertrophic cardiomyopathy: Correlations with the obstructive gradientJACC, 1987
- Noninvasive assessment of valve area in patients with aortic stenosisJACC, 1986
- Determination of the stenotic aortic valve area in adults using Doppler echocardiographyJACC, 1986
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Non-invasive assessment of aortic stenosis by Doppler ultrasound.Heart, 1980