Investigation of Artifacts and Optimization in Proton Resonance Frequency Thermometry Towards Heating Risk Monitoring of Implantable Medical Devices in Magnetic Resonance Imaging
- 18 May 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. 68 (12) , 3638-3646
Abstract
Objective : Artifacts limit the application of proton resonance frequency (PRF) thermometry for on-site, individualized heating evaluations of implantable medical devices such as deep brain stimulation (DBS) for use in magnetic resonance imaging (MRI). Its properties are unclear and the research on how to choose an unaffected measurement region is insufficient. Methods : The properties of PRF signals around the metallic DBS electrode were investigated through simulations and phantom experiments considering electromagnetic interferences from material susceptibility and the radio frequency (RF) interactions. A threshold method on phase difference Δ ϕ was used to define a measurement area to estimate heating at the electrode surface. Its performance was compared to that of the Bayesian magnitude method and probe measurements. Results : The B 0 magnetic field inhomogeneity due to the electrode susceptibility was the main influencing factor on PRF compared to the RF artifact. Δ ϕ around the electrode followed normal distribution but was distorted. Underestimation occurred at places with high temperature rises. The noise was increased and could be well estimated from magnitude images using a modified NEMA method. The Δ ϕ -threshold method based on this knowledge outperformed the Bayesian magnitude method by more than 42% in estimation error of the electrode heating. Conclusion : The findings favor the use of PRF with the proposed approach as a reliable method for electrode heating estimation. Significance : This study clarified the influence of device artifacts and could improve the performance of PRF thermometry for individualized heating assessments of patients with implants under MRI.Keywords
Funding Information
- National Natural Science Foundation of China (51777115, 81830033)
- National Key Research and Development Program of China (2016YFC0105502)
This publication has 44 references indexed in Scilit:
- Artifact properties of carbon nanotube yarn electrode in magnetic resonance imagingJournal of Neural Engineering, 2013
- Confirmation of functional zones within the human subthalamic nucleus: Patterns of connectivity and sub-parcellation using diffusion weighted imagingNeuroImage, 2012
- Assessment of radiofrequency self‐heating around a metallic wire with MR T1‐based thermometryMagnetic Resonance in Medicine, 2011
- Clinical evaluation of MR temperature monitoring of laser‐induced thermotherapy in human liver using the proton‐resonance‐frequency method and predictive models of cell deathJournal of Magnetic Resonance Imaging, 2011
- An agar gel membrane-PDMS hybrid microfluidic device for long term single cell dynamic studyLab on a Chip, 2010
- MR thermometryJournal of Magnetic Resonance Imaging, 2008
- Calculation of MRI‐induced heating of an implanted medical lead wire with an electric field transfer functionJournal of Magnetic Resonance Imaging, 2007
- MRI-based localization of electrophysiological recording sites within the cerebral cortex at single-voxel accuracyNature Methods, 2006
- Bio-heat transfer model of deep brain stimulation-induced temperature changesJournal of Neural Engineering, 2006
- Noninvasive MRI Thermometry with the Proton Resonance Frequency (PRF) Method: In Vivo Results in Human MuscleMagnetic Resonance in Medicine, 1995