The change of response of ionization chambers in the penumbra and transmission regions: impact for IMRT verification
- 8 September 2007
- journal article
- Published by Springer Nature in Medical & Biological Engineering & Computing
- Vol. 46 (4) , 373-380
Abstract
Significant deviations from the expected dose have been reported in the absolute dosimetry validation of an intensity modulated radiation therapy treatment when individual segments are analyzed. However, when full treatment is considered and all segment doses are added together, these discrepancies fade out, leading to overall dose deviations below a 5% action level. This contradictory behavior may be caused by a partial compensation between detector over-responding and under-responding for measurement conditions far from radiation equilibrium. We consider three treatment verification scenarios that may lead to ionization chamber miss-responding, namely: narrow beam irradiation, field penumbra location and multi-leaf collimator transmission contribution. In this work we have analyzed the response of three different ionization chambers with different active volume under these conditions by means of Monte Carlo (MC) simulation methods. Correction factors needed to convert the detector readout into actual dose to water were calculated by inserting the specific detector geometry (carefully modeled) into the simulations. This procedure required extensive use of parallel computing resources in order to achieve the desired level of uncertainty in the final results. The analysis of the simulations shows the relative contribution of each of the three previously mentioned miss-responding scenarios. Additionally, we provide some evidence on dose deviation compensation in multi-segment radiotherapy treatment verification.Keywords
This publication has 25 references indexed in Scilit:
- Portal dose measurements by a 2D arrayPhysica Medica, 2007
- Large efficiency improvements in BEAMnrc using directional bremsstrahlung splittingMedical Physics, 2004
- Ionization chamber‐based reference dosimetry of intensity modulated radiation beamsMedical Physics, 2004
- An EGSnrc Monte Carlo study of the microionization chamber for reference dosimetry of narrow irregular IMRT beamletsMedical Physics, 2004
- The volume effect of detectors in the dosimetry of small fields used in IMRTMedical Physics, 2003
- Dosimetric considerations for validation of a sequential IMRT process with a commercial treatment planning systemPhysics in Medicine & Biology, 2002
- Intensity-modulated radiotherapy: current status and issues of interestInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2001
- Tumor response and salivary function sparing in patients with oropharyngeal squamous cell carcinoma treated with intensity modulated radiation therapy (IMRT) with/without chemotherapy: the mallinckrodt institute of radiology initial resultsInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Use of a new type of radiochromic film, a new parallel-plate micro-chamber, MOSFETs, and TLD 800 microcubes in the dosimetry of small beamsMedical Physics, 1998