Graphic-Output Temperature Data Loggers for Monitoring Vaccine Refrigeration: Implications for Pertussis
Authors:
- Patrick McCollosterPMPatrick McColloster
- Carlos VallbonaCVCarlos Vallbona
- 1 January 2011
- journal article
- research article
- Published by American Public Health Association in American Journal of Public Health
- Vol. 101 (1) , 46-47
Abstract
We conducted the first US study using graphic-output temperature data loggers in quantifying cold chain failure. Fifty-four vaccine refrigerators of a county outpatient health system were studied. Forty-eight percent maintained temperatures of 2°C to 8°C and 24% had protracted periods of temperatures less than 0°C. The correlation between the percentage of refrigerators with freezing temperatures and the pertussis rate for each health region was r = 0.76. The findings suggest that improper vaccine storage may have contributed to recent increases in pertussis rates.Keywords
This publication has 15 references indexed in Scilit:
- Improving vaccination cold chain in the general practice setting.2008
- Freezing temperatures in the vaccine cold chain: A systematic literature reviewVaccine, 2007
- Summary of notifiable diseases--United States, 2004.2006
- Safety and immunogenicity of a combined DTPa–IPV vaccine administered as a booster from 4 years of age: A reviewVaccine, 2006
- Hospitalization for pertussis: profiles and case costs by ageBMC Infectious Diseases, 2005
- Potency of pertussis component in the DTP vaccine – an overview of three decade study in PolandBiologicals, 2004
- Cadena del frío para la conservación de las vacunas en los centros de atención primaria de un área de Madrid: mantenimiento y nivel de conocimientosRevista Espanola de Salud Publica, 2002
- Storage at −3°C for 24 h alters the immunogenicity of pertussis vaccinesVaccine, 2001
- Is the cold chain for vaccines maintained in general practice?BMJ, 1993
- Storage of vaccines in the community: weak link in the cold chain?BMJ, 1992