Allergic contact dermatitis caused by isobornyl acrylate in the Enlite glucose sensor and the Paradigm MiniMed Quick‐set insulin infusion set
- 27 August 2019
- journal article
- research article
- Published by Wiley in Contact Dermatitis
- Vol. 81 (6) , 432-437
Abstract
Background: The FreeStyle Libre glucose sensor has caused many cases of allergic contact dermatitis, and isobornyl acrylate (IBOA) in this sensor has been identified as one of the culprit allergens.Objectives: To report on the presence of IBOA in devices produced by Medtronic, namely, the Enlite sensor and the insulin infusion set Paradigm MiniMed Quick‐set.Patients and Methods: Five patients reacting to the glucose sensor Enlite and/or the insulin infusion set Paradigm MiniMed Quick‐set observed in three clinics (two Belgian and one Swedish) were patch tested with the baseline and other series, as well as with IBOA; four of them also with pieces of adhesive patches from the devices, and two with a thin layer chromatogram of Enlite glucose sensor extracts. Gas chromatography‐mass spectrometry (GC‐MS) analyses were performed.Results: Four patients reacted to IBOA and one to colophonium, a known allergen in Enlite, and three to the adhesive part of the sensor or the insulin infusion set. IBOA was identified in the sensor by GC‐MS, and its presence was indicated in the infusion set.Conclusions: IBOA is a contact allergen in Enlite glucose sensor, and likely also in the infusion set. Therefore, these devices are not suitable alternatives for patients sensitized to the FreeStyle Libre sensor.Keywords
This publication has 11 references indexed in Scilit:
- Allergic contact dermatitis caused by glucose sensors in type 1 diabetes patientsContact Dermatitis, 2019
- Allergic contact dermatitis caused by isobornyl acrylate when using the FreeStyle® LibreJournal of Diabetes Investigation, 2019
- Allergic contact dermatitis caused by isobornyl acrylate in a young diabetic patient using a continous glucose monitoring system (Freestyle Libre)Contact Dermatitis, 2018
- Isobornyl acrylate contained in the insulin patch pump OmniPod as the cause of severe allergic contact dermatitisContact Dermatitis, 2018
- Allergic contact dermatitis and diabetes medical devices: 2 clinical casesContact Dermatitis, 2018
- Allergic contact dermatitis caused by isobornyl acrylate in OmniPod, an innovative tubeless insulin pumpContact Dermatitis, 2018
- Further Evidence of Severe Allergic Contact Dermatitis From Isobornyl Acrylate While Using a Continuous Glucose Monitoring SystemJournal of Diabetes Science and Technology, 2018
- Allergic contact dermatitis caused by isobornyl acrylate in Freestyle® Libre, a newly introduced glucose sensorContact Dermatitis, 2017
- European Society of Contact Dermatitis guideline for diagnostic patch testing – recommendations on best practiceContact Dermatitis, 2015
- Contact allergy to components of glue in insulin pump infusion setsContact Dermatitis, 1995