Effect of the Acoustic Impedance Mismatch at the Bone-Soft Tissue Interface as a Function of Frequency in Transcranial Ultrasound: A Simulation and In Vitro Experimental Study
- 14 December 2020
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 68 (5) , 1653-1663
Abstract
Transcranial Doppler ultrasound (TCD) is a method that uses a hand held low frequency (2-2.5 MHz), pulsed Doppler phased array probe to measure blood velocity within the arteries located inside the brain. The problem with TCD lies in the low ultrasonic energy penetrating inside the brain through the skull which leads to low signal to noise ratio. This is due to several effects including phase aberration, variations in the speed of sound in the skull, scattering, the acoustic impedance mismatch and absorption of the three layer medium constituted by soft tissues, the skull and the brain. The goal of this paper is to study the effect of transmission losses due to the acoustic impedance mismatch on the transmitted energies as a function of frequency. To do so, wave propagation was modelled from the ultrasonic transducer into the brain. This model calculates transmission coefficients inside the brain, leading to a frequency-dependent transmission coefficient for a given skin and bone thickness. This approach was validated experimentally by comparing the analytical results with measurements obtained from a bone phantom plate mimicking the skull. The average position error of the occurrence of the maximum amplitude between the experiment and analytical result was equivalent to a 0.06 mm error on the skin thickness given a fixed bone thickness. Similarity between the experimental and analytical result was also demonstrated by calculating correlation coefficients. The average correlation between the experimental and analytical result came out to be 0.50 for a high frequency probe and 0.78 for a lower frequency probe. Further analysis of the simulation showed that an optimized excitation frequency can be chosen based on skin and bone thicknesses, thereby offering an opportunity to improve the image quality of TCD.Keywords
Funding Information
- International Research Project (IRP) “Laboratoire Franco-Canadien de Recherches Ultrasonores”
- Natural Sciences and Engineering Research Council of Canada (NSRERC) Collaborative Research and Training Experience Program (414043-2012)
This publication has 31 references indexed in Scilit:
- Towards Aberration Correction of Transcranial Ultrasound Using Acoustic Droplet VaporizationUltrasound in Medicine & Biology, 2008
- High-frequency ultrasound of skin involvement in systemic sclerosis reflects oedema, extension and severity in early diseaseRheumatology, 2008
- Local frequency dependence in transcranial ultrasound transmissionPhysics in Medicine & Biology, 2006
- “Ultrasonic stars” for time-reversal focusing using induced cavitation bubblesApplied Physics Letters, 2006
- Image Quality of the Temporal Bone Window Examined by Transcranial Doppler Sonography and Correlation with Postmortem Computed Tomography MeasurementsCerebrovascular Diseases, 2003
- Variations in Facial Skin Thickness and Echogenicity with Site and AgeActa Dermato-Venereologica, 1999
- Time-shift compensation of ultrasonic pulse focus degradation using least-mean-square error estimates of arrival timeThe Journal of the Acoustical Society of America, 1994
- A theory for optimization in the use of acoustic emission transducersThe Journal of the Acoustical Society of America, 1980
- Comprehensive compilation of empirical ultrasonic properties of mammalian tissuesThe Journal of the Acoustical Society of America, 1978
- Acoustical properties of the human skullThe Journal of the Acoustical Society of America, 1978