Simple and portable on-site system for nucleic acid-based detection of Clostridium difficile in stool samples using two columns containing microbeads and loop-mediated isothermal amplification
Authors:
- 3 August 2021
- journal article
- research article
- Published by Springer Nature in Analytical and Bioanalytical Chemistry
- Vol. 414 (1) , 613-621
Abstract
It is challenging to employ nucleic acid-based diagnostics for the in situ detection of Clostridium difficile from complex fecal samples because essential sample preparation and amplification procedures require various experimental resources. In this study, a simple and effective on-site nucleic acid-based detection system was used to detect C. difficile in stool samples. Two columns containing different microbeads, namely, glass and functionalized graphene oxide-coated microbeads, were designed to remove relatively large impurities by filtration and concentrate bacteria, including C. difficile, from stool samples by adsorption. The bacterial nucleic acids were effectively extracted using a small bead beater. The effectiveness of enzyme inhibitors remaining in the sample was efficiently reduced by the direct buffer developed in this study. This sample preparation kit consisting of two simple columns showed better performance in real-time polymerase chain reaction (PCR) and equivalent performance in loop-mediated isothermal amplification (LAMP) than other sample preparation kits, despite 90% simplification of the process. The amplification-ready samples were introduced into two microtubes containing LAMP pre-mixtures (one each for E. coli as an external positive control and C. difficile) by a simple sample loader, which was operated using a syringe. LAMP, which indicates amplification based on color change, was performed at 65 °C in a small water bath. The limit of detection (L.O.D) and analytical sensitivity/specificity of our simple and effective kit were compared with those of a commercial kit. C. difficile in stool samples could be detected within 1 h with 103 cfu/10 mg using LAMP combined simple on-site detection kit.Keywords
This publication has 19 references indexed in Scilit:
- The microfluidic chip module for the detection of murine norovirus in oysters using charge switchable micro-bead beatingBiosensors and Bioelectronics, 2015
- Sensitive and Selective Culture Medium for Detection of Environmental Clostridium difficile Isolates without Requirement for Anaerobic Culture ConditionsJournal of Clinical Microbiology, 2014
- Comparison of culture based methods for the isolation of Clostridium difficile from stool samples in a research settingAnaerobe, 2014
- Graphene and graphene oxide: biofunctionalization and applications in biotechnologyTrends in Biotechnology, 2011
- The role of toxin A and toxin B in Clostridium difficile infectionNature, 2010
- Application of Isothermal Helicase-Dependent Amplification with a Disposable Detection Device in a Simple Sensitive Stool Test for Toxigenic Clostridium difficileThe Journal of Molecular Diagnostics, 2008
- Loop-Mediated Isothermal Amplification Assay for Rapid Detection of Common Strains ofEscherichia coliJournal of Clinical Microbiology, 2008
- Detection of Clostridium difficile Toxin: Comparison of Enzyme Immunoassay Results with Results Obtained by Cytotoxicity AssayJournal of Clinical Microbiology, 2007
- Rapid and Simple Method for Detecting the Toxin B Gene of Clostridium difficile in Stool Specimens by Loop-Mediated Isothermal AmplificationJournal of Clinical Microbiology, 2005
- Amplification of DNA from whole blood.Genome Research, 1994