Relationship Between Anticoagulation Intensity and Thrombotic or Bleeding Outcomes Among Outpatients With Continuous-Flow Left Ventricular Assist Devices

Authors:

Abstract
Background—: We evaluated thrombotic and bleeding outcomes in patients with continuous-flow left ventricular assist devices (CF-LVADs), stratified by anticoagulation intensity. Previous studies of outpatients with CF-LVADs have suggested that target international normalized ratio (INR) values P =0.002). The optimal INR based on weighted mortality of thrombotic and bleeding events was 2.6. Conclusions—: INR is inversely related to thrombotic events occurring outside of the hospital among patients supported with CF-LVADs. INR values <2.0 increase the rate of thrombotic events occurring outside of the hospital among patients supported with CF-LVADs.

This publication has 25 references indexed in Scilit: