In vitro comparison of platinum-iridium and gold tip electrodes: lesion depth in 4 mm, 8 mm, and irrigated-tip radiofrequency ablation catheters
Authors:
- Markus LinhartMLMarkus Linhart
- C. WeissCWC. Weiss
Open Access
- 26 February 2009
- journal article
- research article
- Published by Oxford University Press (OUP) in EP Europace
- Vol. 11 (5) , 565-570
Abstract
We compared a newly developed irrigated gold tip electrode ablation catheter and a gold tip 4 and 8 mm catheter with the corresponding platinum–iridium (Pt) tip catheters in an in vitro setting. In a flow chamber simulating physiological flow conditions, radiofrequency catheter ablation was performed on tissue samples of porcine endomyocardium and liver. Lesion depth, energy and temperature delivery, and popping frequency were determined. Two hundred and fifty-three ablations were conducted. Four and eight millimetre, gold tip electrode catheters produced significantly deeper lesions compared with the Pt tip electrode (liver 4 mm: 4.67 ± 1.7 vs. 2.9 ± 1.0 mm, P < 0.0001; endomyocardium 4 mm: 3.88 ± 1.1 vs. 2.81 ± 0.7 mm, P < 0.001; liver 8 mm: 3.98 ± 1.0 vs. 2.03 ± 1.1 mm, P < 0.001; endomyocardium 8 mm: 4.00 ± 0.9 vs. 3.39 ± 0.8 mm, P < 0.001) and correlated with the amount of energy delivery. Popping frequency was significantly higher in gold tip electrodes. In irrigated tip electrodes, there was no difference in the lesion depth comparing gold with Pt (liver: 5.18 ± 0.7 vs. 5.01 ± 0.7 mm, P = ns; endomyocardium: 4.89 ± 0.7 vs. 4.78 ± 0.8 mm, P = ns). There was a trend towards less popping in the gold tip electrode. Both 4 and 8 mm not-irrigated gold tip catheters produced deeper lesions than the corresponding Pt tip catheter. In irrigated tip catheters, gold and Pt tip material did not show differences in the lesion depth.Keywords
This publication has 18 references indexed in Scilit:
- Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam PopCirculation, 2006
- Gold‐Tip Electrodes—A New “Deep Lesion” Technology for Catheter Ablation? In Vitro Comparison of a Gold Alloy Versus Platinum–Iridium Tip Electrode Ablation CatheterJournal of Cardiovascular Electrophysiology, 2005
- ACC/AHA/ESC Guidelines for the Management of Patients With Supraventricular Arrhythmias—Executive SummaryCirculation, 2003
- Catheter ablation of ventricular tachycardia in patients with structural heart disease using cooled radiofrequency energyJACC, 2000
- Why a Large Tip Electrode Makes a Deeper Radiofrequency Lesion:Journal of Cardiovascular Electrophysiology, 1998
- Comparison of Gold Versus Platinum Electrodes on Myocardial Lesion Size Using Radiofrequency EnergyPacing and Clinical Electrophysiology, 1996
- Radiofrequency Delivery Through a Cooled Catheter Tip Allows the Creation of Larger Endomyocardial Lesions in the Ovine HeartJournal of Cardiovascular Electrophysiology, 1995
- Comparison of In Vivo Tissue Temperature Profile and Lesion Geometry for Radiofrequency Ablation With a Saline-Irrigated Electrode Versus Temperature Control in a Canine Thigh Muscle PreparationCirculation, 1995
- Temperature-guided radiofrequency catheter ablation with very large distal electrodes.Circulation, 1993
- Observations on electrode-tissue interface temperature and effect on electrical impedance during radiofrequency ablation of ventricular myocardium.Circulation, 1990