Abstract
A new, fast, continuous flow technique is described for the simultaneous determination of δ 33 S and δ 34 S using SO masses 48, 49 and 50. Analysis time is ∼5 min/sample with measurement precision and accuracy better than ±0.3‰. This technique, which has been set up using IAEA Ag 2 S standards S‐1, S‐2 and S‐3, allows for the fast determination of mass‐dependent or mass‐independent fractionation (MIF) effects in sulfide, organic sulfur samples and possibly sulfate. Small sample sizes can be analysed directly, without chemical pre‐treatment. Robustness of the technique for natural versus artificial standards was demonstrated by analysis of a Canon Diablo troilite, which gave a δ 33 S of 0.04‰ and a δ 34 S of −0.06‰ compared to the values obtained for S‐1 of 0.07‰ and −0.20‰, respectively. Two pyrite samples from a banded‐iron formation from the 3710 Ma Isua Greenstone Belt were analysed using this technique and yielded MIF (Δ 33 S of 2.45 and 3.31‰) comparable to pyrite previously analysed by secondary ion probe. Copyright © 2004 John Wiley & Sons, Ltd.