A gold film detector for radial profile measurement of plasma density by using a gold neutral beam probe
Authors:
- 25 October 2002
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 73 (11) , 3806-3812
Abstract
A gold film detector was developed in order to measure the plasma density profile by use of the gold neutral beam probe. The detector is a kind of neutral beam detector composed of a rectangular entrance collimator, two sets of grid plates, a gold thin film, a quartz backing, and a microcarbon coated copper plate. The secondary electrons emitted from the gold film were utilized for the neutral beam detection. The yield curves of the secondary electrons were measured as a function of the incident angle and electron energy in both cases of the gold neutral probing beam and ultraviolet ray injection. Time evolution of the plasma line density was measured by adjusting the incident angle and adding a beam chopping method. The result was in good agreement with the line density measured by a microwave interferometer. It was found that this film detector was quite useful in the viewpoint of simultaneous measurements of the electrostatic potential and plasma density profiles.Keywords
This publication has 9 references indexed in Scilit:
- Potential Measurement at the Inner Mirror Throat of the Tandem Mirror GAMMA10Fusion Technology, 2001
- End-Loss Ion Analysis Using an End-Loss Energy Component Analyzer in the Tandem MirrorJournal of the Physical Society of Japan, 1997
- A new method of radial potential profile measurement utilizing magnetic field gradient near a mirror throatReview of Scientific Instruments, 1996
- Observation of loss boundaries and ion flux spectra of end-loss ions in the tandem mirror GAMMA 10Physics of Fluids B: Plasma Physics, 1992
- Enhancement of the thermal barrier depth and two-dimensional potential profiles in GAMMA 10Nuclear Fusion, 1990
- Low-frequency oscillations in the central cell of the TMX tandem mirror experimentThe Physics of Fluids, 1983
- Elmo Bumpy Torus heavy ion probeReview of Scientific Instruments, 1980
- A direct measurement of plasma space potentialNuclear Fusion, 1970
- Plasma Density Measurement by Molecular Ion BreakupReview of Scientific Instruments, 1967