High-dimensional multi-pass flow cytometry via spectrally encoded cellular barcoding
Open Access
- 30 November 2023
- journal article
- research article
- Published by Springer Nature in Nature Biomedical Engineering
- Vol. 8 (3) , 310-324
Abstract
Advances in immunology, immuno-oncology, drug discovery and vaccine development demand improvements in the capabilities of flow cytometry to allow it to measure more protein markers per cell at multiple timepoints. However, the size of panels of fluorophore markers is limited by overlaps in fluorescence-emission spectra, and flow cytometers typically perform cell measurements at one timepoint. Here we describe multi-pass high-dimensional flow cytometry, a method leveraging cellular barcoding via microparticles emitting near-infrared laser light to track and repeatedly measure each cell using more markers and fewer colours. By using live human peripheral blood mononuclear cells, we show that the method enables the time-resolved characterization of the same cells before and after stimulation, their analysis via a 10-marker panel with minimal compensation for spectral spillover and their deep immunophenotyping via a 32-marker panel, where the same cells are analysed in 3 back-to-back cycles with 10–13 markers per cycle, reducing overall spillover and simplifying marker-panel design. Cellular barcoding in flow cytometry extends the utility of the technique for high-dimensional multi-pass single-cell analyses.Keywords
Funding Information
- U.S. Department of Health & Human Services | National Institutes of Health (DP1-EB024242)
- U.S. Department of Health & Human Services | NIH | National Institute of Biomedical Imaging and Bioengineering (R01-EB033155)
- U.S. Department of Health & Human Services | NIH | National Institute of General Medical Sciences (GM139504, R43-GM140527)
This publication has 60 references indexed in Scilit:
- Validation of cell‐based fluorescence assays: Practice guidelines from the ICSH and ICCS – part V – assay performance criteriaCytometry Part B: Clinical Cytometry, 2013
- Optimal Method to Stimulate Cytokine Production and Its Use in Immunotoxicity AssessmentInternational Journal of Environmental Research and Public Health, 2013
- Quantifying spillover spreading for comparing instrument performance and aiding in multicolor panel designCytometry Part A, 2013
- EuroFlow standardization of flow cytometer instrument settings and immunophenotyping protocolsLeukemia, 2012
- EuroFlow antibody panels for standardized n-dimensional flow cytometric immunophenotyping of normal, reactive and malignant leukocytesLeukemia, 2012
- Phospho Flow Cytometry Methods for the Analysis of Kinase Signaling in Cell Lines and Primary Human Blood Samples2010
- Polyethyleneimine Coating Enhances the Cellular Uptake of Mesoporous Silica Nanoparticles and Allows Safe Delivery of siRNA and DNA ConstructsACS Nano, 2009
- Computer simulations of the energy dissipation rate in a fluorescence‐activated cell sorter: Implications to cellsBiotechnology and Bioengineering, 2007
- Fluorescent cell barcoding in flow cytometry allows high-throughput drug screening and signaling profilingNature Methods, 2006
- Differential Modulation of Surface and Intracellular Protein Expression by T Cells after Stimulation in the Presence of Monensin or Brefeldin AClinical and Vaccine Immunology, 2002