THE STERILIZATION OF GLASS AND CALOMEL ELECTRODES BY ULTRAVIOLET IRRADIATION

Authors:
  • RWR. W. Watson
    R. W. Watson
  • MCMary T. Clement
    Mary T. Clement
  • DMD. R. Muirhead
    D. R. Muirhead

Abstract
By strictly defining the procedure for handling and storing the electrodes, contamination was limited to microorganisms from the air. For the purpose of this investigation, therefore, sterilization is defined as the complete inactivation or killing of air-borne bacteria, yeasts, and molds. If the organisms are not present in such a way as to offer mutual protection from the radiation, the total energy at 2537 Å necessary to inactivate the most resistant organism may be assumed to ensure sterilization. Aspergillus niger v. Tiegh. was identified as the most resistant contaminant, and electrostatic sampling showed that clumps containing up to about 40 spores of A. niger Strain 6277 became air-borne. These clumps required approximately 60,000 μw. min. per cm.2 at 2537 Å for inactivation. Used with the accessory procedures, exposure to 680 μw. per cm.2 for two hours was found to provide a safe margin and to ensure sterilization of carefully cleaned and specially mounted electrodes. A satisfactory design for a crevice-free glass electrode sterilizable by ultraviolet light is described. E.m.f. changes in glass–calomel cells sterilized by the recommended procedure do not exceed 0.4 mv., corresponding to 0.006 pH units.