Glucose Clamping Using the Biostator GCIIS

Authors:
  • CVC. A. Verdonk
    C. A. Verdonk
  • RRR. A. Rizza
    R. A. Rizza
  • RWR. E. Westland
    R. E. Westland
  • RNR. L. Nelson
    R. L. Nelson
  • JGJ. E. Gerich
    J. E. Gerich
  • FSF. J. Service
    F. J. Service

Abstract
The Biostator GCIIS was used to clamp circulating glucose at hypoglycemic (42 ± 1 mg/dl), euglycemic (86 ± 2 mg/dl), and hyperglycemic (142 ± 1 mg/dl) levels in normal subjects during a concomitant infusion of insulin (0.1 U/kg/h). Because of limitations in maximal glucosè infusion (1 g/min) from the Biostator, a supplementary infusion of glucose was required to accomplish euglycemic and hyperglycemic clamps. The coefficients of variation of blood glucose were 7.06 ± 1.3 %, 5.9 ± 0.5 %, and 6.1 ± 0.9 for 120 minutes of hypoglycemic, euglycemic, and hyperglycemic clamps, respectively. Despite the occasional interruption of blood flow in the double-lumen catheter and the occasional poor correlation between Biostator and reference glucose method, satisfactory glucose clamps could be maintained for 2 hours.