Abstract
A high‐performance liquid chromatography (HPLC) method developed for analysis of zinc dithiocarbamates was validated and used to perform a survey of disposable medical gloves used in southern Sweden. The gloves were extracted with acetone at room temperature for 10 min by shaking. The extracts were injected into a polyether ether ketone lined HPLC column, and peaks were analysed by a diode‐array detector. The survey shows that of 19 gloves analysed, 10 contained zinc diethyldithiocarbamate (0.070–3.5 mg/g), 3 contained zinc pentamethylenedithiocarbamate (1.0–4.3 mg/g), 4 contained zinc dibutyldithiocarbamate (0.9–1.1 mg/g), and 2 contained 2‐mercaptobenzothiazole (0.005–0.008 mg/g). None of them contained thiurams.