Monopolar Biphasic Stimulator With Discharge Function and Negative Level Shifter for Neuromodulation SoC Integration in Low-Voltage CMOS Process
Authors:
- 7 June 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. 15 (3) , 568-579
Abstract
A 16-channel monopolar biphasic stimulator chip with discharge function for biomedical applications is proposed and designed. To provide monopolar biphasic stimulus currents, the positive (6V) and the negative (-6V) voltage sources are supported to generate the desired current pulses of ±3 mA. The monopolar biphasic stimulator chip was fabricated in a 0.18-μm 1.8-V/3.3-V CMOS process with the common grounded p-type substrate. The overstress and reliability issues on the low-voltage transistors in the stimulator circuits were fully overcome by circuit design. The silicon area of each single channel only occupies 0.08 mm 2 and the output level of stimulus current is up to ±3 mA. By applying the discharge function, safety concern of unbalanced charge in neuro-stimulation can be dealt properly. The residual average dc current is less than 3.42 nA after discharge is activated. Moreover, this chip has also been verified with both in-vitro imitation measurement and in-vivo animal test.Keywords
Funding Information
- Center for Neuromodulation Medical Electronics Systems
- The Featured Areas Research Center Program
- Ministry of Education
- Ministry of Science and Technology (MOST 110-2321-B-009-004, MOST 109-2221-E-009-100-MY3)
This publication has 34 references indexed in Scilit:
- Light-Controlled Biphasic Current Stimulator IC Using CMOS Image Sensors for High-Resolution Retinal Prosthesis and In Vitro Experimental Results With rd1 MouseIEEE Transactions on Biomedical Engineering, 2014
- Simulation Study of the Single-Event Effects Sensitivity in Nanoscale CMOS for Body-Biasing CircuitsIEEE Transactions on Device and Materials Reliability, 2014
- A Fully Intraocular High-Density Self-Calibrating Epiretinal ProsthesisIEEE Transactions on Biomedical Circuits and Systems, 2013
- Adaptive deep brain stimulation in advanced Parkinson diseaseAnnals of Neurology, 2013
- Safety Ensuring Retinal Prosthesis With Precise Charge Balance and Low Power ConsumptionIEEE Transactions on Biomedical Circuits and Systems, 2013
- Comparison of electrical conductivities of various brain phantom gels: Developing a ‘brain gel model’Materials Science and Engineering: C, 2012
- A Tripolar Current-Steering Stimulator ASIC for Field Shaping in Deep Brain StimulationIEEE Transactions on Biomedical Circuits and Systems, 2011
- Monopolar vs. bipolar subretinal stimulation—An in vitro studyJournal of Neuroscience Methods, 2011
- Synchrony in Normal and Focal Epileptic Brain: The Seizure Onset Zone is Functionally DisconnectedJournal of Neurophysiology, 2010
- Neural Stimulation and Recording ElectrodesAnnual Review of Biomedical Engineering, 2008