Investigation of Reproducibility and Error Associated with qPCR Methods using Quantifiler® Duo DNA Quantification Kit*
- 1 September 2010
- journal article
- research article
- Published by Wiley in Journal of Forensic Sciences
- Vol. 55 (5) , 1331-1339
Abstract
Abstract: Reproducibility of quantitative PCR results is dependent on the generation of consistent calibration curves via accurate volume transfers and instrument performance. A review of 14 standard curves, using two different QuantDuo® standard DNA lots, showed variability of cycle threshold values between assays were larger than those of the Internal PCR Control (IPC). This prompted a set of experiments designed to determine the source of variability. Results showed that error introduced during DNA addition to the plate resulted in little variation. A comparison of seven independent series demonstrated cycle threshold variation between dilutions was larger than the variation expected from repeated samples. Modeling the influence of pipette errors on dilution series accuracy indicated that a more rigorous approach to external calibration curve production is required and showed that improvement in calibration curve stability is expected if the pipette conditions are carefully chosen and/or a single validated curve is utilized as the calibrator.Keywords
This publication has 15 references indexed in Scilit:
- Quantifiler™ Observations of Relevance to Forensic Casework*Journal of Forensic Sciences, 2008
- Validation of Reduced‐Scale Reactions for the Quantifiler™ Human DNA KitJournal of Forensic Sciences, 2007
- Developmental validation of a multiplex qPCR assay for assessing the quantity and quality of nuclear DNA in forensic samplesForensic Science International, 2007
- Human Genomic DNA Quantitation System, H‐Quant: Development and Validation for Use in Forensic CaseworkJournal of Forensic Sciences, 2007
- Simultaneous Determination of Total Human and Male DNA Using a Duplex Real‐Time PCR AssayJournal of Forensic Sciences, 2006
- Error propagation in relative real-time reverse transcription polymerase chain reaction quantification models: The balance between accuracy and precisionAnalytical Biochemistry, 2006
- Development of a Human‐Specific Real‐Time PCR Assay for the Simultaneous Quantitation of Total Genomic and Male DNA*Journal of Forensic Sciences, 2006
- Evaluation of Real‐Time PCR Amplification Efficiencies to Detect PCR InhibitorsJournal of Forensic Sciences, 2006
- Analysis of the sensitivity to the systematic error in least-squares regression modelsAnalytica Chimica Acta, 2004
- Corrections for Bias in Regression Estimates After Logarithmic TransformationEcology, 1973