Validation of the TonoVet®rebound tonometer in normal and glaucomatous cats

Authors:

Abstract
Objective To validate intraocular pressure (IOP) readings obtained in cats with the TonoVet®tonometer.Animals studied IOP readings obtained with the TonoVet®were compared to IOP readings determined by manometry and by the Tono‐Pen XL™in 1 normal cat and two glaucomatous cats. TonoVet®and Tono‐Pen XL™readings were also compared in a further six normal and nine glaucomatous cats.Procedures The anterior chambers of both eyes of three anesthetized cats were cannulated and IOP was varied manometrically, first increasing from 5 to 70 mmHg in 5 mmHg increments, then decreasing from 70 to 10 mmHg in 10 mmHg decrements. At each point, two observers obtained three readings each from both eyes, with both the TonoVet®and Tono‐Pen XL™. IOP was measured weekly for 8 weeks with both tonometers in six normal and nine glaucomatous unsedated cats. Data were analyzed by linear regression. Comparisons between tonometers and observers were made by paired studentt‐test.Results The TonoVet®was significantly more accurate than the Tono‐Pen XL™(P = 0.001), correlating much more strongly with manometric IOP. In the clinical setting, the Tono‐Pen XL™underestimated IOP when compared with the TonoVet®.Conclusions Both the TonoVet®and Tono‐Pen XL™provide reproducible IOP measurements in cats; however, the TonoVet®provides readings much closer to the true IOP than the Tono‐Pen XL™. The TonoVet®is superior in accuracy to the Tono‐Pen XL™for the detection of ocular hypertension and/or glaucoma in cats in a clinical setting.