Safety of repetitive nerve stimulation in patients with cardiac implantable electronic devices
- 21 April 2013
- journal article
- research article
- Published by Wiley in Muscle & Nerve
- Vol. 47 (6) , 840-844
Abstract
Introduction: Pacemakers and implantable cardioverter‐defibrillators (ICDs) are vulnerable to inappropriate sensing of electromagnetic interference (EMI), such as from nerve conduction studies. We conducted a prospective study to assess the safety of repetitive nerve stimulation (RNS). Methods: Fourteen patients undergoing insertion of 10 ICDs and 4 pacemakers under general anesthesia received RNS of the median, axillary, and spinal accessory nerves at 2 HZ and 50 HZ. We recorded detection of EMI and whether or not this resulted in an arrhythmia diagnosis or change in pacing output. Results: EMI was visible in 2 ICDs, without spurious tachyarrhythmia detection. EMI was observed in 3 of the 4 pacemakers, which led to pacing inhibition and a pause in 2, both of which were programmed to a unipolar sensing configuration. Conclusions: RNS is safe in patients with ICDs. In pacemaker patients, RNS appears safe during bipolar sensing, but caution is recommended in pacemaker patients with unipolar sensing. Muscle Nerve 47: 840–844, 2013Keywords
This publication has 11 references indexed in Scilit:
- Safety of nerve conduction studies in patients with peripheral intravenous linesMuscle & Nerve, 2010
- Effects of Electromagnetic Interference on Implanted Cardiac Devices and Their ManagementCardiology in Review, 2007
- Safety of nerve conduction studies in patients with implanted cardiac devicesMuscle & Nerve, 2006
- Interference in Implanted Cardiac Devices, Part IIPacing and Clinical Electrophysiology, 2002
- Interference in Implanted Cardiac Devices, Part IPacing and Clinical Electrophysiology, 2002
- Supraventricular Tachycardia‐Ventricular Tachycardia Discrimination Algorithms in Implantable Cardioverter Defibrillators: State‐of‐the‐Art ReviewJournal of Cardiovascular Electrophysiology, 2001
- Interactions Between Electronic Article Surveillance Systems and Implantable Cardioverter-DefibrillatorsCirculation, 1999
- Implantable Defibrillators, Pacemakers, and Electronic Antitheft DevicesNew England Journal of Medicine, 1999
- Interference with Cardiac Pacemakers by Cellular TelephonesNew England Journal of Medicine, 1997
- Inadvertent suppression of a fixed rate ventricular pacemaker using a peripheral nerve stimulatorAnaesthesia, 1993