2D Iterative Image Reconstruction in Dual Head Micro PET with Lookup System Matrix

Authors:

Abstract
The quality of the reconstructed images in small animal studies through micro PET depends largely on the complexity of the model to determine the system matrix. Each system matrix element corresponds to the probability of the detection of an annihilation event inside the field of view (FOV). Monte Carlo methods allow us to simulate physical process not included in the analytical model, such as non collinearity of gamma rays,positron range effect, interaction of incident gamma rays with the scintillator crystal (including attenuation and Compton scatteringeffect), etc. We have evaluated the elements of one of the quadrants of the transaxial central plane. The reflection and rotation symmetries have been use to reduce the system matrix calculation time. We use the 2D‐Maximum Likelihood Expectation Maximization (MLEM) iterative method with Single Slice Rebinning (SSRB) technique to reconstruct a point source and Derenzo‐like phantom images. To evaluate the performance of the Micro PET system, the merit curves and the MOBY mouse phantom reconstruction were obtained.