Webbing and Delamination of Drug Eluting Stent Coatings
Authors:
- 20 October 2015
- journal article
- research article
- Published by Springer Nature in Annals of Biomedical Engineering
- Vol. 44 (2) , 419-431
Abstract
The advancement of the drug-eluting stent technology raises the significant challenge of safe mechanical design of polymer coated stent systems. Experimental images of stent coatings undergoing significant damage during deployment have been reported; such coating damage and delamination can lead to complications such as restenosis and increased thrombogenicity. In the current study a cohesive zone modeling framework is developed to predict coating delamination and buckling due to hinge deformation during stent deployment. Models are then extended to analyze, for the first time, stent-coating damage due to webbing defects. Webbing defects occur when a bond forms between coating layers on adjacent struts, resulting in extensive delamination of the coating from the strut surfaces. The analyzes presented in this paper uncover the mechanical factors that govern webbing induced coating damage. Finally, an experimental fracture test of a commercially available stent coating material is performed and results demonstrate that the high cohesive strength of the coating material will prevent web fracture, resulting in significant coating delamination during stent deployment.Keywords
This publication has 23 references indexed in Scilit:
- FEA Study on the Stress Distributions in the Polymer Coatings of Cardiovascular Drug-Eluting Stent Medical DevicesAnnals of Biomedical Engineering, 2014
- A combined computational and experimental methodology to determine the adhesion properties of stent polymer coatingsComputational Materials Science, 2013
- Anisotropic mode-dependent damage of cortical bone using the extended finite element method (XFEM)Journal of the Mechanical Behavior of Biomedical Materials, 2013
- Cortical bone failure mechanisms during screw pulloutJournal of Biomechanics, 2012
- Changes to Polymer Surface of Drug-Eluting Stents During Balloon ExpansionJAMA, 2012
- Scanning electron microscopic assessment of coating irregularities and their precursors in unexpanded durable polymer‐based drug‐eluting stentsCatheterization and Cardiovascular Interventions, 2011
- Coronary StentsJACC, 2010
- Coronary StentsJACC, 2010
- Computational Investigation of the Delamination of Polymer Coatings During Stent DeploymentAnnals of Biomedical Engineering, 2010
- Evaluation of drug‐eluting stents' coating durability—Clinical and regulatory implicationsJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2009