Intra‐oral ignition of monopolar diathermy during transnasal humidified rapid‐insufflation ventilatory exchange ( THRIVE )
Authors:
- 20 March 2017
- journal article
- Published by Wiley in Anaesthesia
- Vol. 72 (6) , 781-783
Abstract
Summary: We present the case of unanticipated airway ignition during hard palate biopsy. Transnasal humidified rapid‐insufflation ventilatory exchange ( THRIVE ) and monopolar diathermy were utilised for the procedure, during which an arc arose from the diathermy tip to a titanium implant, causing a brief ignition on the monopolar diathermy grip. This case highlights the need for maintained awareness of fire risk when using diathermy in the presence of THRIVE during airway surgery.Keywords