Low torque levels can initiate a removal of the passivation layer and cause fretting in modular hip stems
- 19 July 2014
- journal article
- Published by IOP Publishing in Medical Engineering & Physics
- Vol. 36 (9) , 1140-1146
Abstract
Taper connections of modular hip prostheses are at risk of fretting and corrosion, which can result in reduced implant survival. The purpose of this study was to identify the minimum torque required to initiate a removal of the passivation layer at the taper interface as a function of assembly force and axial load. Titanium stems and cobalt-chromium heads were assembled with peak impaction forces of 4.5 kN or 6.0 kN and then mounted on a materials testing machine whilst immersed in Ringer's solution. The stems were subjected to a static axial load (1 kN or 3 kN) along the taper axis. After a period of equilibration, a torque ramp from 0 to 15 Nm was manually applied and the galvanic potential was continuously recorded. Prostheses assembled with a force of 6 kN required a significantly higher torque to start a removal of the passivation layer compared to those assembled with 4.5 kN (7.23±0.55 Nm vs. 3.92±0.97 Nm, p=0.029). No influence of the axial load on the fretting behaviour was found (p=0.486). The torque levels, which were demonstrated to initiate surface damage under either assembly force, can be readily reached during activities of daily living. The damage will be intensified in situations of large weight and high activity of the patient or malpositioning of the prosthesis.Keywords
This publication has 29 references indexed in Scilit:
- Modular titanium alloy neck adapter failures in hip replacement - failure mode analysis and influence of implant materialBMC Musculoskeletal Disorders, 2010
- The influence of the centre of rotation on implant survival using a modular stem hip prosthesisInternational Orthopaedics, 2008
- Fretting crevice corrosion of stainless steel stem–CoCr femoral head connections: Comparisons of materials, initial moisture, and offset lengthJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2008
- Pseudotumours associated with metal-on-metal hip resurfacingsThe Journal of Bone and Joint Surgery. British volume, 2008
- Die Bedeutung der Setzkraft für die Sicherheit einer Konuskopplung von Hüftstiel und keramischem Prothesenkopf / Relevance of the insertion force for the taper lock reliability of a hip stem and a ceramic femoral headBiomedical Engineering / Biomedizinische Technik, 2006
- Effect of impact assembly on the fretting corrosion of modular hip tapersJournal of Orthopaedic Research, 2005
- In vitro corrosion testing of modular hip tapersJournal of Biomedical Materials Research Part B: Applied Biomaterials, 2002
- The Morse Taper Junction in Modular Revision Hip Replacement – A Biomechanical and Retrieval Analysis - Die konische Steckverbindung in der modularen Revisionsendoprothetik der Hüfte – Vergleich eines explantierten Prothesenschaftes zur In-vitro-TestungBiomedical Engineering / Biomedizinische Technik, 2000
- In vivo corrosion of modular hip prosthesis components in mixed and similar metal combinations. The effect of crevice, stress, motion, and alloy couplingJournal of Biomedical Materials Research, 1993
- Interface corrosion of a modular head total hip prosthesisThe Journal of Arthroplasty, 1992