MR imaging of the cervical spine: assessment of image quality with parallel imaging compared to non-accelerated MR measurements
- 27 October 2006
- journal article
- research article
- Published by Springer Nature in European Radiology
- Vol. 17 (5) , 1147-1155
Abstract
To compare the quality of cervical spine MR images obtained by parallel imaging [generalized autocalibrating partially parallel acquisition (GRAPPA)] with those of non-accelerated imaging, we conducted both phantom studies and examinations of ten volunteers at 1.5Tesla with a dedicated 12-element coil system and a head-spine-neck coil combination. Acquisitions included axial T2-weighted (T2w) images with both methods, and sagittal T2w and T1w images in vivo with the latter coil combination. Non-accelerated MRI with two averages and GRAPPA (acceleration factor 2) with two averages (GRAPPA/2AV, time reduction of approximately 50%) and four averages (GRAPPA/4AV) were compared. In the phantom, the signal-to-noise ratio of the GRAPPA/2AV was lower than that of the other two settings. In vivo, the image inhomogeneity (non-uniformity, NU) was significantly higher in T2w GRAPPA/2AV than in both other settings, and in T1w GRAPPA/2AV compared to GRAPPA/4AV. Subjectively, the delineation of anatomical structures was sufficient in all sequences. Only the spinal cord was considered to be better delineable on the non-accelerated T1w sequence compared to GRAPPA/2AV. In part, GRAPPA/4AV performed better than the other settings. The subjective image noise was lowest with GRAPPA/4AV. In cervical spine MRI, the examination time can be reduced by nearly 42% with GRAPPA, while preserving sufficient image quality.Keywords
This publication has 22 references indexed in Scilit:
- Comparison of In Vitro and In Vivo MRI of the Spine Using Parallel ImagingAmerican Journal of Roentgenology, 2004
- Rapid Musculoskeletal Magnetic Resonance Imaging Using Integrated Parallel Acquisition Techniques (IPAT) - Initial ExperiencesRöFo - Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren, 2003
- Transmit SENSEMagnetic Resonance in Medicine, 2002
- Generalized autocalibrating partially parallel acquisitions (GRAPPA)Magnetic Resonance in Medicine, 2002
- Coil Sensitivity Encoding in MR ImagingAmerican Journal of Roentgenology, 2002
- 2D sense for faster 3D MRIMagnetic Resonance Materials in Physics, Biology and Medicine, 2002
- Recent advances in image reconstruction, coil sensitivity calibration, and coil array design for SMASH and generalized parallel MRIMagnetic Resonance Materials in Physics, Biology and Medicine, 2002
- A comparison of two methods for measuring the signal to noise ratio on MR imagesPhysics in Medicine & Biology, 1999
- SENSE: Sensitivity encoding for fast MRIMagnetic Resonance in Medicine, 1999
- Correction of intensity nonuniformity in MR images of any orientationMagnetic Resonance Imaging, 1993