Review and evaluation of MRI nonuniformity corrections for brain tumor response measurements
Authors:
- John J. HeineJHJohn J. Heine
- Alan B. CantorACAlan B. Cantor
- Hongbo LinHLHongbo Lin
- Lynn M. FletcherLFLynn M. Fletcher
- Laurence P. ClarkeLCLaurence P. Clarke
- 1 September 1998
- journal article
- review article
- Published by Wiley in Medical Physics
- Vol. 25 (9) , 1655-1666
Abstract
Current MRI nonuniformity correction techniques are reviewed and investigated. Many approaches are used to remedy this artifact, but it is not clear which method is the most appropriate in a given situation, as the applications have been with different MRI coils and different clinical applications. In this work four widely used nonuniformity correction techniques are investigated in order to assess the effect on tumor response measurements (change in tumor volume over time): a phantom correction method, an image smoothing technique, homomorphic filtering, and surface fitting approach. Six brain tumor cases with baseline and follow‐up MRIs after treatment with varying degrees of difficulty of segmentation were analyzed without and with each of the nonuniformity corrections. Different methods give significantly different correction images, indicating that rf nonuniformity correction is not yet well understood. No improvement in tumor segmentation or in tumor growth/shrinkage assessment was achieved using any of the evaluated corrections.Keywords
This publication has 54 references indexed in Scilit:
- Statistical approach to segmentation of single-channel cerebral MR imagesIEEE Transactions on Medical Imaging, 1997
- Sources of intensity nonuniformity in spin echo images at 1.5 TMagnetic Resonance in Medicine, 1994
- Radiotherapy response of cerebral metastases quantified by serial MR imagingJournal of Neuro-Oncology, 1994
- Accurate intensity correction for endorectal surface coil MR imaging of the prostateIEEE Transactions on Nuclear Science, 1993
- Long term study of random noise and signal uniformity in spin-echo brain imagingMedical Physics, 1993
- Correction of intensity variations in MR images for computer-aided tissue classificationIEEE Transactions on Medical Imaging, 1993
- A technique for accurate magnetic resonance imaging in the presence of field inhomogeneitiesIEEE Transactions on Medical Imaging, 1992
- Eddy current disruption: Effect on nuclear magnetic resonance coil impedance and power lossMedical Physics, 1989
- The information content of MR imagesIEEE Transactions on Medical Imaging, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986