Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the NIM, China, and the BIPM in low-energy x-rays

Authors:
  • DBD T Burns
    D T Burns
  • CKC Kessler
    C Kessler
  • WJWu Jinjie
    Wu Jinjie
  • WPWang Peiwei
    Wang Peiwei

Abstract
A key comparison has been made between the air-kerma standards of the NIM and the BIPM in the low-energy x-ray range. The results show the standards to be in agreement at the level of the combined standard uncertainty of 3.7 parts in 10 3 . The results are analysed and presented in terms of degrees of equivalence, suitable for entry in the BIPM key comparison database. KEY WORDS FOR SEARCH Radiation dosimetry, key comparison, low-energy x-rays, air kerma Main text To reach the main text of this paper, click on Final Report . Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/ . The final report has been peer-reviewed and approved for publication by the CCRI, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).