Simulated evaluation of a non-Luer safety connector system for use in neuraxial procedures
Authors:
Open Access
- 23 November 2011
- journal article
- research article
- Published by Elsevier in British Journal of Anaesthesia
- Vol. 108 (1) , 134-139
Abstract
Spinal syringes, needles, and other devices with connectors that will not also connect with Luer devices could substantially reduce wrong-route drug administration errors. This study aimed to evaluate a newly designed non-Luer safety connector system for neuraxial procedures in terms of clinical acceptability and cross-connectivity with conventional Luer devices. A non-Luer safety connector system (BD UniVia-6 Safety Connector system), which included non-Luer spinal needles, syringes, and blunt fill filter needles, was evaluated in a prospective, simulated use, randomized study. Clinicians evaluated the acceptability and performance characteristics using a normal saline injectate on an artificial back model. Forty-nine clinicians participated in the study. For 93 of 98 spinal injection procedures, clinicians agreed that overall, the safety system was clinically acceptable (94.9%; 95% lower bound 89.6%). Forty-eight clinicians (98%; 95% lower bound: 90.7%) agreed that the safety system prevented or reduced the risk of misconnection between a conventional syringe and a safety spinal needle. A lower proportion (63.3%; 95% lower bound: 50.5%) agreed that the safety system would prevent or reduce the risk of misconnections between a safety syringe filled with medication intended for spinal injection and an i.v. Luer device. Our study demonstrated that study clinicians found the safety system to be acceptable with minimal impact on technique in a simulated-use setting. The non-Luer system also appeared to decrease the risk of administration of i.v. medications into the intrathecal space. Further modifications will be required to completely eliminate the risk of administering intrathecal medication i.v. and evaluation in a clinical setting will help determine the true impact of this non-Luer system on patient safety.Keywords
This publication has 11 references indexed in Scilit:
- Accidental Intrathecal Injection of Aminophylline in Spinal AnesthesiaAnesthesiology, 2011
- Fatal consequences of a simple mistake: How can a patient be saved from inadvertent intrathecal vincristine?Clinical Neurology and Neurosurgery, 2011
- Replacing Luer connectors: still work in progressAnaesthesia, 2010
- A simulation‐based evaluation of two proposed alternatives to Luer devices for use in neuraxial anaesthesia*Anaesthesia, 2010
- The quest to eliminate intrathecal vincristine errors: a 40-year journeyQuality and Safety in Health Care, 2010
- Accidental Intrathecal Administration of VincristineAmerican Journal of Forensic Medicine & Pathology, 2010
- Intrathecal VincristineJournal of Pediatric Hematology/Oncology, 2009
- Medication errors in anesthesia: an 8-year retrospective analysis at an urban university hospitalJournal of Anesthesia, 2008
- Inadvertent intrathecal injection of tramadolBritish Journal of Anaesthesia, 2003
- Convulsions and Refractory Ventricular Fibrillation after Intrathecal Injection of a Massive Dose of Tranexamic AcidAnesthesiology, 2003