Integrated probe for sweat analysis
Authors:
- 1 January 1988
- journal article
- research article
- Published by Wiley in Journal of Clinical Laboratory Analysis
- Vol. 2 (1) , 16-18
Abstract
Solid‐state integrated probe based on ion‐sensitive, field‐effect transistors has been used for determination of sodium chloride in sweat for diagnosis of cystic fibrosis. This probe can yield useful clinical data faster and with higher reliability than the conventional titration test. A battery‐powered instrument that converts the differential current into milliequivalents of sodium chloride has been constructed. It is possible to perform the diagnosis without inducing sweating with pilocarpine. It is suggested that similar probes with multisensing capability can be used for analysis of other surface body fluids.Keywords