2D Images Recorded With a Single-Sided Magnetic Particle Imaging Scanner
- 17 December 2015
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 35 (4) , 1056-1065
Abstract
Magnetic Particle Imaging is a new medical imaging modality, which detects superparamagnetic iron oxide nanoparticles. The particles are excited by magnetic fields. Most scanners have a tube-like measurement field and therefore, both the field of view and the object size are limited. A single-sided scanner has the advantage that the object is not limited in size, only the penetration depth is limited. A single-sided scanner prototype for 1D imaging has been presented in 2009. Simulations have been published for a 2D single-sided scanner and first 1D measurements have been carried out. In this paper, the first 2D single-sided scanner prototype is presented and the first calibration-based reconstruction results of measured 2D phantoms are shown. The field free point is moved on a Lissajous trajectory inside a 30 ×30 mm 2 area. Images of phantoms with a maximal distance of 10 mm perpendicular to the scanner surface have been reconstructed. Different cylindrically shaped holes of phantoms have been filled with 6.28 μl undiluted Resovist. After the measurement and image reconstruction of the phantoms, particle volumes could be distinguished with a distance of 2 mm and 6 mm in vertical and horizontal direction, respectively.Keywords
Funding Information
- Bundesministerium für Forschung, Technologie und Raumfahrt
- European Union and Schleswig-Holstein (122-10-004)
This publication has 25 references indexed in Scilit:
- Magnetic particle imaging scanner with 10-kHz drive-field frequencyBiomedical Engineering / Biomedizinische Technik, 2013
- On the formulation of the image reconstruction problem in magnetic particle imagingBiomedical Engineering / Biomedizinische Technik, 2013
- Magnetic particle imaging: Introduction to imaging and hardware realizationZeitschrift für Medizinische Physik, 2012
- An x-space magnetic particle imaging scannerReview of Scientific Instruments, 2012
- Weighted iterative reconstruction for magnetic particle imagingPhysics in Medicine & Biology, 2010
- Signal encoding in magnetic particle imaging: properties of the system functionBMC Medical Imaging, 2009
- Experimental results on fast 2D-encoded magnetic particle imagingPhysics in Medicine & Biology, 2008
- Cancer Statistics, 2008CA: A Cancer Journal for Clinicians, 2008
- Tomographic imaging using the nonlinear response of magnetic particlesNature, 2005
- A concept for the clinical implementation of sentinel lymph node biopsy in patients with breast carcinoma with special regard to quality assuranceCancer, 2005