Dual-Frequency Intravascular Sonothrombolysis: An In Vitro Study
- 9 August 2021
- journal article
- research article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 68 (12) , 3599-3607
Abstract
Thrombo-occlusive disease is one of the leading causes of death worldwide. There has been active research on safe and effective thrombolysis in preclinical and clinical studies. Recently, the dual-frequency transcutaneous sonothrombolysis with contrast agents [microbubbles (MBs)] has been reported to be more efficient in trigging the acoustic cavitation, which leads to a higher lysis rate. Therefore, there is increasing interest in applying dual-frequency technique for more significant efficacy improvement in intravascular sonothrombolysis since a miniaturized intravascular ultrasound transducer typically has a limited power output to fully harness cavitation effects. In this work, we demonstrated this efficacy enhancement by developing a new broadband intravascular transducer and testing dual-frequency sonothromblysis in vitro . A broadband intravascular transducer with a center frequency of 750 kHz and a footprint size of 1.4 mm was designed, fabricated, and characterized. The measured −6-dB fractional bandwidth is 68.1%, and the peak negative pressure is 1.5 MPa under the driving voltage of 80 V pp . By keeping one frequency component at 750 kHz, the second frequency component was selected from 450 to 650 kHz with an interval of 50 kHz. The in vitro sonothrombolysis tests were conducted with a flow model and the results indicated that the MB-mediated, dual-frequency (750+500 kHz) sonothrombolysis yields an 85% higher lysis rate compared with the single-frequency treatment, and the lysis rate of dual-frequency sonothrombolysis increases with the difference between the two frequency components. These findings suggest a dual-frequency excitation technique for more efficient intravascular sonothrombolysis than conventional single-frequency excitationKeywords
Funding Information
- National Institutes of Health (R01HL141967, R21EB027304)
This publication has 37 references indexed in Scilit:
- Safety of Thrombolysis in Acute Ischemic Stroke: A Review of Complications, Risk Factors, and Newer TechnologiesThe Neurohospitalist, 2011
- Image-Guided Tumor Ablation: Emerging Technologies and Future DirectionsSeminars in Interventional Radiology, 2010
- Venous ThromboembolismAmerican Journal of Preventive Medicine, 2010
- Low-dose Systemic Thrombolytic Therapy for Deep Vein Thrombosis in Pediatric PatientsJournal of Pediatric Hematology/Oncology, 2010
- Noninvasive Thrombolysis Using Pulsed Ultrasound Cavitation Therapy – HistotripsyUltrasound in Medicine & Biology, 2009
- Single-transducer dual-frequency ultrasound generation to enhance acoustic cavitationUltrasonics Sonochemistry, 2009
- Theoretical and experimental validation of a dual-frequency excitation method for spatial control of cavitationPhysics in Medicine & Biology, 2005
- Ultrasound-Enhanced Systemic Thrombolysis for Acute Ischemic StrokeNew England Journal of Medicine, 2004
- The Epidemiology of Venous ThromboembolismCirculation, 2003
- Albumin Microbubble Echo-Contrast Material as an Enhancer for Ultrasound Accelerated ThrombolysisCirculation, 1995