Abstract
In this paper, an on-line solid-phase extraction capillary electrophoresis-mass spectrometry (SPE-CE-MS) method is described for purification, preconcentration, separation and characterization of endogenous microRNA (miRNA) and their post-transcriptional modifications in serum. First, analysis by CE-MS was optimized using a standard mixture of hsa-miR-21-5p (miR-21-5p) and hsa-let-7g-5p (let-7g-5p). For SPE-CE-MS, a commercial silicon carbide (SiC) resin was used to prepare the microcartridges. Under the optimized conditions using standards, microcartridge lifetime (>25 analyses) and repeatability (2.8% RSD for migration times; 4.4 and 6.4% RSD for miR-21-5p and let-7g-5p peak areas, respectively) were good, the method was linear between 25-100 nmol•L-1 and limit of detection (LOD) was around 10 nmol•L-1 (50 times lower than by CE-MS). In order to analyze human serum samples, an off-line sample pretreatment based on phenol:chloroform:isoamyl alcohol (PCA) extraction was necessary prior to SPE-CE-MS. The potential of the SPE-CE-MS method to screen for B-cell chronic lymphocytic leukemia (CLL) was demonstrated analyzing serum samples from healthy controls and patients. MicroRNAs were only detected in CLL patients, specifically miR-21-5p and a 23-nucleotide long 5’-phosphorylated miRNA with 3’-uridylation (iso-miR-16-5p).
Funding Information
  • Ministerio de Econom?a y Competitividad (CTQ2014-56777-R)
  • Ministerio de Educaci?n, Cultura y Deporte

This publication has 57 references indexed in Scilit: