Purpose: The aim of this study was to measure the increase in temperature in dental implants during the intake of hot beverages in vivo. Materials and Methods: Eight successfully osseointegrated implants in 7 subjects were examined. Each subject was asked to drink the same volume of hot beverage. While drinking, temperature changes were recorded via 3 embedded thermocouples placed (i) in the implant’s internal space, (ii) at the implant-abutment interface, and (iii) at the abutment. All thermocouples were linked to a computer and analyzed with appropriate software. Results: The maximum temperatures were 47.3°C at the abutment, 45.6°C at the implant’s internal space, and 44.6°C at the implant-abutment interface. A linear correlation was found between the temperatures measured (i) at the implant abutment and in the implant’s internal space, and (ii) at the abutment and at the abutment-implant interface. Conclusions: Further clinical studies are required to determine whether the habitual consumption of hot food and beverages may be considered a risk factor in the success of implant-supported prostheses.Keywords
This publication has 16 references indexed in Scilit:
- Frailty Approach for the Analysis of Clustered Failure Time Observations in Dental ResearchJournal of Dental Research, 2005
- Basic principles of thermal dosimetry and thermal thresholds for tissue damage from hyperthermiaInternational Journal of Hyperthermia, 2003
- Thermal changes observed at implant sites during internal and external irrigationClinical Oral Implants Research, 2002
- Heat shock‐induced necrosis and apoptosis in osteoblastsJournal of Orthopaedic Research, 1999
- Heat generated during preparation of titanium implants of the ITI® Dental Implant System: an in vitro studyClinical Oral Implants Research, 1995
- Temperatures during drilling for the placement of implants using the osseointegration techniqueJournal of Oral and Maxillofacial Surgery, 1986
- Heat caused by drilling cortical bone: Temperature measuredin vivoin patients and animalsActa Orthopaedica, 1984
- Temperature threshold levels for heat-induced bone tissue injury: A vital-microscopic study in the rabbitThe Journal of Prosthetic Dentistry, 1983
- Thermal injury to boneInternational Journal of Oral Surgery, 1982
- Osseointegrated Titanium Implants:Requirements for Ensuring a Long-Lasting, Direct Bone-to-Implant Anchorage in ManActa Orthopaedica, 1981