Technique for Removal of Intramedullary Nails When There Is Failure of the Proximal Extraction Device: A Report of Three Cases
Authors:
- Bruce E. HeckBHBruce E. Heck
- Nabil A. EbraheimNENabil A. Ebraheim
- 1 February 1997
- journal article
- case report
- Published by Wolters Kluwer Health in Journal of Orthopaedic Trauma
- Vol. 11 (2) , 130-132
Abstract
When attempting to remove an intramedullary nail, the extraction device can fail. This can occur due to stripping of the proximal threads or, more rarely, due to entrapped broken metal devices (impacted cap screws; broken introducers, extractors, or taps). This latter situation is particularly problematic if access to the intramedullary portion of the nail and proximal threads is not possible. A technique for implant removal utilizing a high-speed drill is described in this report.Keywords
This publication has 7 references indexed in Scilit:
- Use of a Tap Extractor to Remove a Broken Tap within a Grosse-Kempf Intrameduallary NailThe Journal of Trauma: Injury, Infection, and Critical Care, 1995
- Case Report: Techniques for the Removal of Broken Small-Diameter Tibial Nails: A Report of Two CasesJournal of Orthopaedic Trauma, 1995
- Complicated Removal of the Distal Locking Device of Brooker-WillsClinical Orthopaedics and Related Research, 1993
- Difficulties during removal of fluted femoral intramedullary rods.Journal of Bone and Joint Surgery, 1989
- Broken intramedullary nails.Journal of Bone and Joint Surgery, 1988
- A Technique for Removing a Fractured Interlocking Tibial NailJournal of Orthopaedic Trauma, 1988
- Alternative Method for Removing an Impacted AO Intramedullary NailClinical Orthopaedics and Related Research, 1986