Operative intra-aortic balloon rupture

Authors:

Abstract
Rupture of an intra-aortic balloon (IAB), inserted to assist in weaning from cardiopulmonary bypass, occurred during attempted intra-aortic administration of protamine in a 71-year-old male who later died. Intra-aortic balloon counterpulsation (IABC) is most commonly utilized to assist in weaning from cardiopulmonary bypass (CPB) or to augment low-flow states following CPB. In-hospital survival following IAB insertion in these circumstances in patients with coronary artery disease is approximately 60 per cent. Patients with valvular disease have a lower in-hospital survival rate (50 per cent). Complications of IABC are usually of vascular or infectious origin. Balloon rupture is a rare though potentially lethal complication. The effects of balloon rupture may be compounded by the use of helium as a driving gas to inflate the balloon. Intra-aortic administration of protamine has not been shown to be superior to peripheral administration and should be avoided if an IAB is in place.

This publication has 19 references indexed in Scilit: