A Computational Study of the Effects of Inlet Guide Vanes on the Performance of a Centrifugal Blood Pump
Authors:
- 18 June 2002
- journal article
- research article
- Published by Wiley in Artificial Organs
- Vol. 26 (6) , 534-542
Abstract
This article presents computational studies on the effects of inlet guide vanes (IGVs) on the flow pattern and shear stress in a centrifugal blood pump. The effect of IGVs is to introduce a pre‐swirl to fluid particles entering the impeller with the intention that the fluid particles will travel along the blade profile. Currently, most commercial centrifugal blood pumps employ straight radial impeller blades that are not hydrodynamically ideal for a good flow pattern within the blade passage. Flow separation and formation of vortices within the blade passage are believed to increase the degree of hemolysis and thrombosis. These are causes for blood clotting that will lead to malfunctioning of ventricular assist devices. Four IGVs of different geometrical profiles have been numerically investigated using a commercial software program CFX‐Tascflow. The pump is operated at 2,000 rpm, and the results revealed that the relative flow patterns in the blade passage have been dramatically altered. The size of the vortices was reduced, and the pressure contours indicated a gradual rise from the impeller leading edge to the trailing edge. However, inclusion of IGV causes a drop in the pressure head generated. Higher frictional losses are incurred as fluid particles passed through the IGV. In addition, the IGV modifies the inlet velocity triangles, and this also contributes to a drop in the pressure head generated that is consistent with Euler's pump theory. The change in the flow patterns and the gradual variation of the pressure contours have led to lower shear stress within the blade passages as compared to the case without IGVs.Keywords
This publication has 9 references indexed in Scilit:
- Fluid Engineering Aspect for Development of the Centrifugal Blood Pump with Magnetically Suspended Impeller2000
- Experimental and Computational Studies of the Relative Flow Field in a Centrifugal Blood PumpCritical Reviews in Biomedical Engineering, 2000
- Development of Design Methods of a Centrifugal Blood Pump with In Vitro Tests, Flow Visualization, and Computational Fluid Dynamics: Results inHemolysis TestsArtificial Organs, 1998
- A Magnetically Suspended and Hydrostatically Stabilized Centrifugal Blood PumpArtificial Organs, 1996
- Prototype Continuous Flow Ventricular Assist Device Supported on Magnetic BearingsArtificial Organs, 1996
- A Quantitative Visualization Study of Flow in a Scaled‐up Model of a Centrifugal Blood PumpArtificial Organs, 1996
- Recent Studies of the Centrifugal Blood Pump with a Magnetically Suspended ImpellerArtificial Organs, 1995
- A Fluid Dynamic Analysis of a Rotary Blood Pump for Design ImprovementArtificial Organs, 1993
- Low haemolysis pulsatile impeller pump: design concepts and experimental resultsJournal of Biomedical Engineering, 1989